B. Braun Space Infusion Pump System

Occlusion Alarms Despite Free-Flowing Line

Asset Type

Infusion Pump

Manufacturer

B. Braun

Model

Space Infusion Pump System

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting B. Braun Space Infusion Pumps that generate occlusion alarms even though the IV line appears to be flowing freely.

In many cases, the cause is not a true occlusion but rather issues such as tubing setup, cassette seating, pressure from clamps, or sensor detection problems. This guide focuses on external checks first, which are the most common causes before assuming an internal pump issue.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the occlusion alarm message and location

Inspect the IV tubing path through the pump

Confirm the correct B. Braun administration set is being used

Check for partially closed clamps

Inspect the line for subtle kinks or compression

Verify the cassette door is fully closed and latched

Prime or flush the line

Check for downstream resistance at the catheter site

Test with a known-good administration set

Power cycle the pump

If the Problem Persists

If occlusion alarms continue after verifying:

Then the issue may involve:

At this point the device should be:

Recognizing when external causes have been ruled out is an important part of proper troubleshooting.

Clinical Use Tip

Do not troubleshoot infusion pump occlusion alarms while connected to an active patient whenever possible.

If an alarm persists:

Patient therapy should never be interrupted while diagnosing equipment issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nursing reported the infusion pump repeatedly alarming for occlusion even though the IV line appeared to be flowing normally.”

Cause

What was observed during troubleshooting.

Example:
“Tubing set was partially misaligned in the pump channel, causing incorrect pressure detection by the occlusion sensor.”

Resolution

What action was taken.

Example:
“Reseated the administration set and verified correct routing. Pump tested and infusion ran normally without additional occlusion alarms.”

Helpful Details to Include (If Known)

Final Thought

Occlusion alarms are designed to protect patients from interrupted therapy, but they can sometimes be triggered by simple setup issues. By starting with tubing placement, clamps, and basic setup checks, Clinical Engineering can quickly rule out common causes before escalating to repair. Clear troubleshooting and proper documentation ensure both patient safety and efficient equipment management.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide