BD Alaris Infusion Pump

Upstream Occlusion / “Check IV Set” Alarms

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where a BD Alaris infusion pump generates Upstream Occlusion or “Check IV Set” alarms. These alarms indicate that the pump is detecting resistance or lack of fluid flow before the pumping mechanism, typically somewhere between the fluid container and the pump.

Many of these alarms are caused by simple external issues such as closed clamps, kinked tubing, improperly installed administration sets, or empty fluid containers. This guide focuses on identifying those common causes before assuming a device failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Alarm Message

Check the IV Fluid Container

Inspect Roller Clamps and Slide Clamps

Check for Kinks or Pinched Tubing

Verify Proper IV Set Installation

Check the Drip Chamber

Inspect the Spike Connection

Try a Different IV Set

Test with a Known Good Setup

If the Problem Persists

If upstream occlusion or “Check IV Set” alarms continue after verifying clamps, tubing, fluid source, and IV set installation, the issue may involve internal pump components such as:

At this point the device should be:

Escalating the device for further evaluation after ruling out external causes is appropriate Clinical Engineering troubleshooting.

Clinical Use Tip

Never troubleshoot infusion pump flow issues while the device is actively connected to a patient.

If a pump alarm occurs during therapy:

Maintaining uninterrupted therapy and patient safety must always come first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nursing reported repeated ‘Upstream Occlusion / Check IV Set’ alarms during infusion setup.”

Cause

What was observed during troubleshooting.

Example:
“Inspection revealed the roller clamp on the IV tubing was partially closed, restricting flow to the pump.”

Resolution

What action was taken.

Example:
“Opened the roller clamp fully and verified normal infusion operation. Alarm cleared and pump returned to service.”

Helpful Details to Include (If Known)

Final Thought

Upstream occlusion alarms on infusion pumps are commonly caused by simple setup issues such as clamps, tubing kinks, or improperly installed IV sets. Clinical Engineering troubleshooting should always begin with these easily verified external factors before assuming device failure. When the basic causes are ruled out, promptly removing the device from service and documenting findings protects both patient safety and equipment reliability.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide