GE Healthcare LOGIQ E Series

Image Artifacts or Poor Image Quality

Asset Type

Ultrasound System

Manufacturer

GE Healthcare

Model

LOGIQ E Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where the GE LOGIQ E Series ultrasound system produces poor image quality or visible artifacts. These issues may appear as:

Many image quality problems are caused by probe issues, cable damage, improper settings, or poor probe contact rather than internal system failure.

Step-by-Step Troubleshooting

1. Confirm the Reported Issue

This helps determine if the issue is probe-related, configuration-related, or system-wide.

2. Inspect the Probe for Physical Damage

Damaged probes commonly produce image artifacts or signal dropouts.

3. Clean the Probe Face

Contamination on the probe surface can distort the ultrasound beam and cause image artifacts.

4. Verify Proper Probe Connection

A partially seated connector can cause signal loss or intermittent imaging artifacts.

5. Test with a Different Probe

This is one of the most effective isolation steps.

6. Check System Imaging Settings

Improper settings can sometimes appear as image degradation or artifacts.

7. Perform a System Reboot

After rebooting, reconnect the probe and test again.

8. Test Multiple Probe Ports

9. Check for Environmental Interference

Interference can occasionally introduce noise or artifacts in ultrasound imaging.

If the Problem Persists

If image artifacts remain after verifying probe condition, connections, and system settings, common external causes have been ruled out.

The issue may involve:

At this point, the device should be:

Recognizing when the issue likely requires internal repair is an important part of proper troubleshooting.

Clinical Use Tip

Never troubleshoot an ultrasound imaging problem while the device is actively being used on a patient.

Move the patient to another ultrasound system before performing troubleshooting steps. This ensures patient care is not delayed and avoids introducing risk during diagnostic imaging.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“User reported that the ultrasound image appeared grainy with horizontal lines during abdominal scans.”

Cause

What was observed during troubleshooting.

Example:
“Inspection revealed damage to the probe cable near the strain relief causing intermittent signal loss.”

Resolution

What action was taken.

Example:
“Probe removed from service and replaced with a functional probe. Damaged probe sent for vendor repair.”

Helpful Details to Include

Final Thought

Ultrasound image quality issues are often caused by probes, connections, or configuration rather than internal system failure. A structured troubleshooting approach that begins with external inspections and simple tests helps isolate the true cause efficiently. Proper documentation and knowing when to remove equipment from service protects both patients and equipment reliability.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide