DeVilbiss 525 Series

Preventive Maintenance Procedure

On this page

Asset Type

Oxygen Concentrator

Manufacturer

DeVilbiss

Model

525 Series

Models Covered

525DS, 525DS-Q, 525KS, 525KS-LT, and 525PS.

Manual Reference

Drive DeVilbiss Healthcare 5-Liter Oxygen Concentrator Service Manual LT-2023 and applicable 525 Series Instruction Guide A-525DS/SE-525K4.

Manufacturer Recommendation / AEM Status

The schedule assumes approximately 5,000 operating hours per year in a normal, clean environment. The provider determines and documents an appropriate interval for actual conditions. Changes to manufacturer-listed activities or intervals belong under the facility's approved AEM program, risk assessment, maintenance history, operating environment, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Step-by-Step PM Procedure

1. Verify Identification and Configuration

Action: Record asset tag, serial, catalog model, rating, hours, auxiliary-port configuration, filters, accessories, location, and environment. Confirm applicable documentation and parts.

Expected outcome: CMMS identity and model-specific requirements are correct.

2. Review History and Environment

Action: Review purity, alarm, flow, pressure, heat, vibration/noise, power, filter, compressor, and prior-result history. Note excessive dust, lint, heat, humidity, or heavy usage.

Expected outcome: Conditions requiring corrective work or a shorter interval are identified.

3. Prepare Safely

Action: Provide alternate prescribed oxygen before removing an in-use unit. Switch off, unplug, remove accessories, and use a ventilated test area free of smoking, flame, sparks, oil, grease, and oxygen-enriched materials.

Expected outcome: Patient support is maintained and fire/electrical hazards are controlled.

4. Inspect Exterior and Mobility

Action: Inspect cabinet, handle, casters, filter door, vents, exhaust, outlet, flowmeter/knob, auxiliary port/cap, labels, hour meter, and fasteners for damage, blockage, contamination, looseness, liquid, or heat.

Expected outcome: The unit is intact, stable, unobstructed, and correctly labeled.

5. Inspect Power Components

Action: Inspect the cord, plug or rear IEC connection, polarization, insulation, strain relief, switch, circuit breaker, fuse holder when fitted, and indicators.

Expected outcome: Power components and controls are undamaged and safe.

6. Complete Due Cleaning and Reprocessing

Action: Clean the cabinet and vents by the instruction guide. Do not spray, saturate, allow liquid into openings, use petroleum solvents, or apply oil/grease. For patient change, qualified personnel shall replace listed non-reusable accessories and perform the complete manufacturer cleaning/disinfection, contact-time, residue-removal, drying, inspection, and purity-verification process.

Expected outcome: The unit is clean, dry, and fully reprocessed when applicable.

7. Inspect External Filters

Action: Determine whether cabinet filter 303DZ-605 is fitted; units manufactured since October 2020 may not have one. Inspect at every PM. Wash with warm water/dishwashing detergent, rinse, dry completely, and reinstall, or replace if damaged. Never operate without a required filter or with it damp. Clean filter-door vents.

Expected outcome: Required filtration is clean, dry, intact, and correctly installed.

8. Inspect Intake Bacteria Filter

Action: Inspect intake bacteria filter MC44D-605 during every PM and replace as necessary using the service procedure.

Expected outcome: The intake filter and filter door are serviceable and restored.

9. Complete Qualified Internal/Compressor Inspection When Due

Action: Qualified personnel only: unplug before cabinet removal. Inspect dust, tubing, wiring, abrasion, heat, corrosion, leaks, mounts, compressor, and rotary-valve area. During compressor service inspect/replace final bacteria filter PV5LD-651 and the applicable compressor filter. Install the final filter with its IN fitting toward the flowmeter. Restore the complete cabinet before power.

Expected outcome: Internal components are clean, secure, correctly filtered, and enclosed.

10. Verify Startup Alert and Indicators

Action: Connect rated mains and switch on. Confirm Service Required, Low Oxygen, and Normal Oxygen indicators illuminate briefly and the audible alert sounds. Allow up to 15 minutes for Normal Oxygen indication.

Expected outcome: Startup self-check passes and Normal Oxygen replaces the startup indication.

11. Verify Power-Failure Alarm

Action: With the unit operating, unplug it by the manufacturer method. Confirm the audible power-failure alert, then reconnect and restore operation.

Expected outcome: Mains loss produces an audible alert.

12. Verify Flow and Low-Flow Alarm

Action: Exercise the flow control through the intended 0.5-5 LPM range and compare selected points with a calibrated flow analyzer. Confirm 5 LPM maximum. After several minutes of operation, use the service-manual method to set flow to zero and confirm the red Service Required light and audible alarm. Restore flow immediately.

Expected outcome: Flow control and maximum delivery pass the approved criterion, and low-flow alarm activates.

13. Verify Oxygen Concentration

Action: Calibrate the analyzer under conditions similar to the concentrator location. Set 5 LPM, connect at the outlet, run at least 20 minutes, and record concentration. Apply the exact model limit: 525DS/525DS-Q: 87-96%; 525KS/525KS-LT/525PS: 93% ±3%.

Expected outcome: Stabilized concentration meets the exact model specification; a green OSD light alone is not a substitute for measurement.

14. Verify Outlet Pressure

Action: Using the calibrated pressure assembly and service-manual setup, measure and record outlet pressure. Do not improvise an outlet obstruction.

Expected outcome: Outlet pressure is 8.5 ±0.5 psig (58.6 ±3.5 kPa).

15. Evaluate Additional Alarm Tests When Required

Action: Routine PM documents startup, power-failure, low-flow, and oxygen-status behavior. Perform service-level overheating or compressor-failure alarm tests only when specifically required and only by the exact qualified procedure; both deliberately simulate faults.

Expected outcome: Required alarm tests pass and no protective function remains disconnected.

16. Perform Electrical Safety Testing

Action: Verify the label. LT-2023 identifies the 525 Series as Class II, double-insulated, so protective-earth resistance is normally not applicable. Perform facility-required Class II touch/enclosure leakage, polarity, and cord tests. Do not impose a grounded-device test solely because another PM uses one.

Expected outcome: Applicable Class II safety tests meet facility requirements.

17. Restore and Return to Service

Action: Restore filters, doors, cabinet, caps, outlet, flow setting, cord, and approved accessories. Confirm normal sound/vibration, unobstructed airflow, stable 5 LPM operation, Normal Oxygen indication, and no alarm, odor, or abnormal heat. Document readings and parts.

Expected outcome: The concentrator is fully assembled, within specification, and ready for service.

Electrical Safety Testing

LT-2023 classifies the 525 Series as Class II, double-insulated equipment. Protective-earth resistance is normally not applicable; verify the exact label. Record the analyzer/standard, classification, cord/polarity inspection, touch/enclosure leakage result, facility limit, and pass/fail status. The reviewed manual does not provide one routine PM leakage-current limit, so use current facility/NFPA 99 criteria for the exact configuration.

Pass / Fail Criteria

Pass

Remove From Service

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide