Hillrom CareAssist

Preventive Maintenance Procedure

On this page

Asset Type

Hospital Bed

Manufacturer

Hillrom

Model

CareAssist

Manual Reference

Hill-Rom CareAssist and CareAssist ES Bed Service Manual, Product No. P1170, MAN330 REV 4.

Manufacturer Recommendation / AEM Status

The manufacturer advises annual bed service. Batteries are replaced every 3 years and bed-exit tape switches every 5 years by facility-authorized maintenance personnel. The annual checklist covers overall condition, structure, welds, labels, hardware, headboard/footboard, AC cord, leakage current, pendant, columns, actuators, sleep deck, CPR, gas springs, articulation, battery, siderails, foot extension, pivot points, casters, fifth wheel, brakes/steering, accessories, and fitted scale, Bed Exit, SideCom, and SafeView features. The manual states that only facility-authorized personnel should service the bed.

Use the current serial- and configuration-specific service manual. If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Step-by-Step PM Procedure

1. Verify Identity and Configuration

Action: Record asset number, CareAssist model, serial number, product number, revision, voltage, and fitted scale, Bed Exit, SideCom, SafeView, and fifth-wheel options.

Expected outcome: The correct requirements and manual apply.

2. Review History and Prepare the Bed

Action: Review prior failures and repairs. Remove the bed from clinical use, remove accessories/load as needed, unplug when directed, and follow facility infection-control and lockout practices.

Expected outcome: Known issues are addressed and service hazards controlled.

3. Inspect Overall Condition

Action: Inspect the bed structure, frame, base, welds, symmetry, corrosion, impact damage, paint, labels, and loose or missing hardware.

Expected outcome: The frame/base are not twisted and all safety-related components are intact and readable.

4. Inspect Headboard and Footboard

Action: Check condition, attachment, fit, and safe removal/reinstallation.

Expected outcome: Both boards are intact, secure, and safe.

5. Inspect the AC Power Cord and Plug

Action: Inspect the full cord, strain relief, and one-piece molded plug for cuts, cracking, loose blades, discoloration, overheating, arcing, or exposed conductors.

Expected outcome: No damage; replace a defective cord assembly.

6. Perform Leakage-Current Testing

Action: Connect the bed through the safety tester, confirm the bed is supplied by it, and measure without operating bed functions.

Expected outcome: Leakage current is less than 100 µA; investigate the cord and power supply if outside specification.

7. Inspect and Test the Patient Pendant

Action: Inspect the connector and cable, reconnect securely, and hold each button several seconds.

Expected outcome: Each button commands the correct function and motion remains continuous without intermittency.

8. Inspect and Test Hi-Low Columns

Action: Check column fasteners and retaining hardware, then fully raise and lower the bed. Verify the Bed Not Down indicator behavior.

Expected outcome: Smooth, quiet motion without friction or overload indication; indicator extinguishes in low position.

9. Inspect and Test the Head Section

Action: Inspect actuator pins and retaining clips, fully raise/lower the head section, and inspect the plastic sleep deck and drive system for binding or damage.

Expected outcome: Intact hardware and smooth, quiet motion.

10. Test CPR Releases and Gas Springs

Action: Inspect both CPR handles, cables, mechanism, screws, and gas-spring attachments/shafts. Test CPR from both sides and confirm the mechanism relocks with Head Up.

Expected outcome: Controlled lowering from high to intermediate and then low, no oil on shafts, and positive relock.

11. Inspect and Test Knee, Foot, and Automatic Contour

Action: Inspect actuator pins/retainers; fully exercise knee and foot sections. Confirm automatic contour raises the knee to mid-height with head elevation and lowers it as the head returns low.

Expected outcome: Correct coordinated motion without abnormal noise, friction, or overload.

12. Test Dining Chair and Trendelenburg Functions

Action: With functions unlocked, operate Dining Chair, return the deck flat, and test Trendelenburg and reverse Trendelenburg.

Expected outcome: All sections reach the commanded positions smoothly and quietly.

13. Check Battery Operation and Age

Action: Unplug the bed and operate all functions from caregiver controls. Inspect battery dates.

Expected outcome: Battery operation is normal; charge/replace as needed and replace batteries when over 3 years old.

14. Test Scale and Bed Exit, If Fitted

Action: Calibrate the scale using the exact manual procedure and listed weights. Function-test all Bed Exit modes/alarms and review the five-year tape-switch interval.

Expected outcome: Scale calibration completes, Bed Exit works properly, and due tape switches are replaced.

15. Inspect Siderails and Caregiver Controls

Action: Check head/foot siderails for bending or twisting, verify audible latch engagement, inspect covered fasteners and cable routing, and test caregiver controls.

Expected outcome: Siderails latch securely and cables/controls are undamaged and functional.

16. Inspect Foot Extension and Pivot Points

Action: Operate the foot extension through its positions and confirm locking when the lever is released. Exercise articulations and investigate noise; lubricate only as directed.

Expected outcome: Secure locking and quiet, unrestricted articulation.

17. Inspect Casters, Fifth Wheel, Brakes, and Steering

Action: Inspect tread/cuts/wear, confirm fifth-wheel motion if fitted, apply brakes and try to move the bed, then verify steering.

Expected outcome: Wheels are serviceable, brakes prevent movement, and steering works correctly.

18. Test Brake-Not-Set Alarm

Action: Connect AC power, place brakes in neutral, and confirm the alarm sounds; set the brakes and confirm it stops.

Expected outcome: Alarm behavior matches brake position.

19. Inspect Accessories and Optional Communication/Alerts

Action: Check fitted accessories. Inspect/test the SideCom junction box, features, cable, and connector pins and test SafeView alerts when fitted.

Expected outcome: Every installed feature is secure and functional.

20. Restore, Final-Test, and Document

Action: Correct defects with approved parts, restore covers/accessories, complete a final function/readiness check, label the bed, and document measurements, replacements, and disposition.

Expected outcome: The bed is complete and safe to return or clearly removed from service.

Electrical Safety Testing

Leakage Current

Touch/chassis leakage measured in microamps:

Manufacturer-listed limit: < 100 µA

Result =

Pass / Fail =

Ground Resistance

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different.

Result =

Use the current service manual and facility procedure for tester configuration and any protective-earth criterion.

Pass / Fail Criteria

PM Pass Criteria

Remove From Service Criteria

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Sources