On this page
Asset Type
Manufacturer
Model
Configurations Covered
Standard P8000, P8000 with scale, P8000 with Steering Plus, and legacy P8000 with IntelliDrive Transport System, as fitted.
Manual References
Hillrom Transport, Procedural, and Specialty Stretchers Instructions for Use 144385 Rev 12 and the applicable Hillrom Service Manual 144386.
Manufacturer Recommendation / AEM Status
Hillrom service guidance recommends annual preventive maintenance and testing. The exact current service-manual revision and serial/configuration-specific checklist control the work. The manufacturer checklist covers overall condition, casters, braking/steering, pedals, hydraulic functions, siderails, sleep deck, mattress, pivots, gas springs, handles, accessories, and installed options.
The Instructions for Use lists an expected service life of approximately ten years under normal use when recommended PM is performed; condition, support, history, and policy still determine disposition. Inspect sooner after collision, fall, fluid intrusion, unusual movement, heavy use, or a reported brake, siderail, hydraulic, steering, or transport-system problem. Any changed interval belongs under the facility's approved AEM program.
What This PM Procedure Covers
- Identification, configuration, frame, welds, fasteners, labels, mattress, deck, bumpers, and accessories
- Casters, brake/steer controls, Steering Plus, and legacy IntelliDrive when fitted
- Hydraulic hi-low, Trendelenburg/reverse Trendelenburg, pedals, leaks, and position holding
- Siderails, back-section gas spring, CPR release, headrest, knee crank, cassette lift, and Automatic Contour
- Scale and powered-option electrical checks when fitted
Required Tools and Equipment
- Current Hillrom 144385 Instructions for Use and serial-appropriate 144386 Service Manual/checklist
- Appropriate hand/torque tools and manufacturer-approved lubricants
- Manufacturer-approved cleaning/disinfection supplies and PPE
- Straightedge or measurement tool for Steering Plus checks when applicable
- Stopwatch and inclinometer when facility-required
- Calibrated test weights for an installed scale, used by the current Hillrom procedure
- Electrical safety analyzer for mains-powered options when required
Step-by-Step PM Procedure
1. Verify Identification and Configuration
Action: Record asset tag, serial, product/revision code, manufacture date, department, safe working load, and options. Identify Steering Plus, scale, Active Brake, IntelliDrive, knee articulation, cassette lift, oxygen storage, IV poles, handles, and accessories. Confirm the correct service manual.
Expected outcome: The stretcher matches CMMS and only applicable tests/specifications are selected.
2. Review History and Prepare Safely
Action: Review brake, caster, steering, hydraulic, siderail, mattress, scale, transport, collision, cleaning, and electrical complaints. Remove from patient use, unload attachments/storage, set the brake, and disconnect powered options before inspection.
Expected outcome: Known problems are identified and patient, pinch, movement, electrical, and load hazards are controlled.
3. Inspect Structure and Labels
Action: Inspect frame, base, welds, deck, shrouds, bumpers, mounting points, hardware, labels, and finish for dents, twist, cracks, corrosion, missing clips/fasteners, loose parts, damaged paint, or asymmetry.
Expected outcome: Structure and welds are intact; hardware and labels are complete; no unsafe damage exists.
4. Inspect Mattress and Deck
Action: Remove and inspect mattress cover, seams, closure, retention, and accessible core for punctures, cuts, tears, stains, fluid ingress, delamination, compression, or lost impermeability. Inspect the deck for sharp edges, contamination, damage, and secure mounting.
Expected outcome: Mattress is intact, impermeable, cleanable, and retained on an undamaged deck.
5. Inspect and Clean Casters
Action: Inspect/clean tires, forks, bearings, stems, fasteners, and swivels. Check for debris, cuts, flat spots, wear, separation, wobble, noise, or binding while rolling and rotating each caster.
Expected outcome: Casters roll/swivel freely, track correctly, and have secure serviceable tread and hardware.
6. Verify Brake, Neutral, and Steer
Action: Operate controls from every available position. In brake, push in multiple directions and confirm the unloaded stretcher does not roll. In neutral, confirm all casters roll/swivel. In steer, confirm straight tracking and correct steer-caster/Steering Plus engagement. Inspect pedals, pads, linkages, and hardware.
Expected outcome: Brake holds, neutral permits free movement, steer tracks, and controls latch without excessive force or lost motion.
7. Verify Steering Plus When Fitted
Action: Inspect center wheel, tread, pivot, linkage, pads, and fasteners. Confirm deployment in steer and retraction in neutral/brake. Verify over-center linkage and floor clearance by the current manual. Older service guidance lists approximately 3.5 inches (89 mm) clearance with brake set; the current serial-specific manual controls.
Expected outcome: Steering Plus deploys, retracts, tracks, and clears the floor without wear or interference.
8. Inspect Pump and Control Pedals
Action: Inspect hi-low, Trendelenburg, reverse Trendelenburg, brake/steer, and option pedals for wear, missing pads, bending, loose pivots, linkage interference, and labeling. Operate each from all locations.
Expected outcome: Pedals move/return freely and command only the intended function.
9. Verify Hi-Low Hydraulics
Action: Unloaded, pump from minimum to maximum height. Older Hillrom guidance specifies no more than 28 pump strokes; verify the current serial-specific limit. Confirm even raising, smooth pressure-compensated lowering under an approved safe test condition, correct stops, and position holding.
Expected outcome: The stretcher raises within the applicable criterion, lowers smoothly, and holds without creeping, binding, or abnormal noise.
10. Verify Trendelenburg and Reverse Trendelenburg
Action: Operate both ranges unloaded. Confirm controlled movement, correct response, holding, and return to level while keeping personnel/equipment clear.
Expected outcome: Both functions move smoothly, hold securely, and return level without drift, binding, or asymmetry.
11. Inspect Hydraulic Components
Action: Inspect cylinders, rods, hoses, fittings, connections, pumps, and surrounding areas for oil, leaks, scratches, nicks, dents, corrosion, looseness, or contamination. Follow the manual for adjustment, bleeding, or repair.
Expected outcome: Components are dry, intact, secure, and hold all tested positions.
12. Inspect Siderails and Latches
Action: Cycle each siderail. Inspect tubes, pivots, handles, hardware, stops, and spacing. Apply controlled force after latching. Confirm positive independent latching, smooth release, and no excessive play, cracking, bending, or pinch hazard.
Expected outcome: Both siderails travel smoothly, latch fully, resist unintended release, and remain structurally sound.
13. Inspect Back Section, Gas Spring, and CPR Release
Action: Cycle the back section. At full extension inspect the gas-spring rod for scratches, nicks, dents, leakage, or corrosion. Confirm seated jam nut, secure pivots/bolts, and no drift. While controlling the section, test each CPR release and inspect cable, nuts, handle, and relatching.
Expected outcome: Back section moves/holds safely; CPR release works without binding or uncontrolled drop; gas spring/hardware pass.
14. Inspect Other Articulations
Action: Test the headrest, knee/thigh crank, cassette lift, and Automatic Contour when fitted. Check motion, stops, locks, spring tension, pivots, lubrication, and binding. Confirm Automatic Contour raises the knee section with the head and lowers appropriately on return.
Expected outcome: Installed articulations operate, lock, and return correctly without excessive play, force, or wear.
15. Inspect Handles and Accessories
Action: Inspect push handles, IV poles/sockets, oxygen-tank storage, drainage hooks, trays, restraint points, and installed accessories. Verify locks, stops, retainers, labels, hardware, and stowage.
Expected outcome: Accessories are secure, functional, correctly labeled, and do not impair movement or stability.
16. Verify Scale When Fitted
Action: Inspect controls, display, cables, mounts, and battery. Zero with approved linens/accessories arranged normally and verify at selected points using calibrated weights and the current Hillrom procedure. The IFU lists accuracy as 1% or 2.2 lb (1 kg), whichever is greater; confirm the exact configuration specification.
Expected outcome: Scale zero, controls, display, battery, and measurements meet the applicable specification.
17. Verify Legacy IntelliDrive When Fitted
Action: Inspect drive wheel, chains, push-handle controls, cables, batteries, manual-mode switch, indicators, and hardware. Confirm each handle commands forward/reverse, movement is controlled without loud or metal-to-metal noise, wheel tread is serviceable, and the wheel lowers in steer and raises in brake/neutral. Address a flashing service indicator by the current service procedure. Later checklists call for battery replacement when older than three years; verify the applicable revision.
Expected outcome: IntelliDrive controls, stops, deploy/retract action, wheel, battery, chains, sound, and indicators pass.
18. Perform Configuration-Appropriate Electrical Safety
Action: A fully manual/hydraulic P8000 does not require AC protective-earth or leakage testing. For installed scale, Active Brake, IntelliDrive charger/system, or other electrical options, inspect power/battery components and perform only the tests required by that option's manual and facility standard. Record measured values and limits; do not borrow limits from another configuration.
Expected outcome: All electrical inspections and option-specific tests meet current criteria, or are correctly documented N/A.
19. Lubricate and Adjust as Specified
Action: Lubricate caster, linkage, crank, and pivot points only where the current service manual directs, using specified materials/amounts. Make adjustments only by approved procedures and remove excess lubricant from patient-contact and walking surfaces.
Expected outcome: Specified points are serviced without contamination, excess lubricant, or unauthorized adjustment.
20. Restore and Return to Service
Action: Restore mattress, accessories, batteries, covers, fasteners, cords, storage items, and configuration. Confirm brake set, siderails latched, stretcher at safe low height, controls neutral, and powered options ready. Repeat a brief operational check, clean, label, and document.
Expected outcome: The P8000 is clean, stable, complete, functional, and safe for clinical service.
Electrical Safety Testing
Select tests by installed configuration. Standard hydraulic units with no mains-connected equipment are normally N/A. For powered options, document the option, exact Hillrom requirement, analyzer/standard, power-component inspection, protective-earth resistance and/or touch/enclosure leakage where applicable, facility limits, and pass/fail. Do not omit required testing on powered options or impose AC tests on a fully manual unit.
Pass / Fail Criteria
Pass
- Serial/configuration, safe working load, labels, options, and manuals are verified
- Frame, welds, hardware, deck, mattress, bumpers, shrouds, and accessories are intact
- Casters, brake, neutral, steer, and Steering Plus when fitted pass
- Hi-low raises evenly within the current criterion, lowers smoothly, and holds without leaks/drift
- Trendelenburg, siderails, back section, CPR release, and all installed articulations pass
- Scale and IntelliDrive meet applicable criteria when fitted
- Required electrical-safety and option tests pass
- No collision damage, corrosion, sharp edge, instability, fluid ingress, oil leak, unusual noise, binding, or unintended movement remains
Remove From Service
- A structural, mattress, fastener, caster, bumper, handle, IV/oxygen holder, or accessory safety defect exists
- Brake/steering is unreliable or a caster/wheel binds, wobbles, or is excessively worn
- A siderail or articulation cannot latch, hold, release, or move safely
- Hydraulic leakage, uneven lift, excessive pump strokes, drift, uncontrolled lowering, or positioning failure exists
- CPR release, gas spring, scale, IntelliDrive, Active Brake, control, battery, or service indicator fails
- Required electrical safety fails or authorized repair cannot be completed
Copy/Paste CMMS Work Order Procedure
Use the current Hillrom service manual for the exact serial and configuration. Confirm older numerical criteria before use if a current revision specifies a different value.
Sources
- Hillrom Transport, Procedural, and Specialty Stretchers Instructions for Use, 144385 Rev 12
- Hillrom Transport, Procedural, and Specialty Stretchers Service Manual, 144386 - use the current serial/configuration-specific revision supplied by Hillrom/Baxter.