Invacare Perfecto2

Preventive Maintenance Procedure

On this page

Asset Type

Oxygen Concentrator

Manufacturer

Invacare

Model

Perfecto2

Models Covered

IRC5PO2AW (SensO2) and IRC5PAW (non-SensO2).

Manual Reference

Invacare Perfecto2 Service Manual, Part No. 1154245.

Manufacturer Recommendation / AEM Status

The reviewed manual requires PM at a minimum of one year in service and more often where excessive dust or soot is present. Verify the exact model, market, and current manual revision because later Perfecto2-family publications may differ. Changes to the applicable manufacturer schedule belong under the facility's approved AEM program.

What This PM Procedure Covers

Required Tools and Equipment

Step-by-Step PM Procedure

1. Verify Identification and Configuration

Action: Record asset tag, serial, exact model, rating, hours, SensO2/non-SensO2 configuration, accessories, location, and applicable manual revision.

Expected outcome: The CMMS identity, interval, and acceptance criteria are correct.

2. Review History and Environment

Action: Review complaints, alarms, oxygen trends, flow/pressure problems, filters, compressor work, prior results, hours, and exposure to dust, soot, heat, or restricted ventilation.

Expected outcome: Adverse trends and need for corrective work or a shorter interval are identified.

3. Prepare Safely

Action: Provide alternate prescribed oxygen before removing an in-use unit. Switch off, unplug, remove patient accessories, and use a clean, ventilated area away from smoking, flame, sparks, oil, grease, and oxygen-enriched materials.

Expected outcome: Patient support is maintained and service hazards are controlled.

4. Inspect the Exterior

Action: Inspect cabinet, handle, feet/casters, seams, fasteners, vents, exhaust, outlet, flowmeter/knob, hour meter, labels, switch, breaker, cord, plug, and strain relief for damage, contamination, blockage, looseness, heat, liquid, or modification.

Expected outcome: Enclosure, airflow, controls, power components, and markings are serviceable.

5. Clean Cabinet and Service Cabinet Filter

Action: Clean without spraying or saturating the unit. Vacuum the cabinet filter or wash in warm soapy water, rinse thoroughly, dry completely, and reinstall. Replace if damaged. Never operate without the filter or with it wet.

Expected outcome: Cabinet and vents are clean; filter is dry, intact, and correctly installed.

6. Check the Outlet HEPA Filter

Action: With the cabinet assembled, operate at maximum flow and attach the specified 50-foot cannula. If the flow indicator fluctuates, investigate and replace the outlet HEPA filter as needed by the qualified service procedure.

Expected outcome: Flow remains stable with no indication of outlet-filter restriction.

7. Check Compressor Inlet Filter and Due Internal Maintenance

Action: Qualified personnel only: unplug before cabinet removal. Inspect/replace the compressor inlet filter as needed. When due or indicated, clean the heat exchanger without deforming it, inspect/service the exhaust muffler, and evaluate the compressor top end. Inspect tubing, wiring, mounts, and connections. Restore the cabinet before power.

Expected outcome: Filters and cooling paths are serviceable and internal components are secure.

8. Verify Startup and General Operation

Action: Place the assembled unit at least 3 inches (7.6 cm) from obstructions. Connect directly to rated mains without an extension cord and switch on. Confirm normal compressor, flowmeter, and indicator operation with no abnormal noise, vibration, odor, or heat.

Expected outcome: The unit starts and operates normally with unobstructed ventilation.

9. Verify Flow

Action: Exercise the control and compare indicated flow with a calibrated analyzer at selected points. Record at least maximum flow. Do not set above 5 L/min.

Expected outcome: Flow adjusts smoothly and meets the approved criterion through 0.5-5 L/min.

10. Verify Low-Flow Indication

Action: With patient accessories removed, briefly reduce flow into the low-flow region by the applicable manual method. Confirm the rapid audible indication occurs within 0-0.5 L/min. Restore normal flow immediately.

Expected outcome: Low-flow warning activates in range and clears after flow is restored.

11. Verify Oxygen Concentration

Action: Calibrate the oxygen analyzer, operate at test flow under normal conditions, and allow approximately 30 minutes to stabilize. Record concentration and verify 87-95.6% oxygen at 0.5-5 L/min. Check IRC5PO2AW SensO2 models annually/at scheduled PM/between patients; check IRC5PAW non-SensO2 units every 180 days. For SensO2, confirm indication agrees with measurement: nominal green above 85% ±2%, yellow from 73% ±3% to 85% ±2%, and red below 73% ±3%. Do not deliberately degrade concentration during routine PM to simulate service-level alarms.

Expected outcome: Concentration meets specification and SensO2 indication, when fitted, agrees.

12. Verify Outlet Pressure

Action: Use the calibrated gauge and approved service-manual setup to measure and record maximum outlet pressure. Do not improvise an obstruction.

Expected outcome: Outlet pressure is 5.0 ±0.5 psi (34.5 ±3.45 kPa).

13. Verify Power-Loss Alarm

Action: With the unit operating, disconnect mains power. Confirm the audible alarm sounds immediately or within 30 seconds. Reconnect and restore operation.

Expected outcome: Mains loss produces the required audible warning.

14. Additional Alarm or Leak Tests When Required

Action: Low/high-pressure, timeout, oxygen-sensor, and pressure-equalization-valve tests deliberately simulate internal faults. Perform only when required and only by qualified personnel following the exact service manual. If pneumatic parts were disturbed, complete the manual leak test.

Expected outcome: Required functions pass and nothing remains disconnected or leaking.

15. Perform Electrical Safety Testing

Action: Verify exact rating-label construction. Inspect power components and perform facility-required testing for that configuration. Record protective-earth resistance only if a protective-earth conductor exists; otherwise mark N/A. Record touch/enclosure leakage and the facility limit.

Expected outcome: Visual and applicable electrical-safety tests pass.

16. Restore and Return to Service

Action: Restore filters, panels, fasteners, caps, accessories, and settings. Confirm stable 5 L/min operation, compliant concentration/pressure, normal indicators, no alarm, and no abnormal sound, vibration, odor, or heat. Document readings and parts.

Expected outcome: The unit is complete, within specification, and ready for service.

Electrical Safety Testing

The reviewed manual does not provide a universal routine leakage-current or protective-earth limit. Verify the exact rating label and apply the facility criterion for its construction. Document analyzer/standard, cord and controls, equipment construction, ground resistance or N/A, touch/enclosure leakage, facility limit, result, and pass/fail.

Pass / Fail Criteria

Pass

Remove From Service

Copy/Paste CMMS Work Order Procedure

Use the exact model's current manufacturer manual, facility policy, and approved AEM program to determine the final interval and return-to-service requirements.

Source

Invacare Perfecto2 Service Manual, Part No. 1154245

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide