What This Page Explains
This page covers:
- - What transcutaneous pacing is
- - Pacing rate
- - Pacing current
- - Fixed versus demand pacing
- - ECG sensing
- - R-wave detection
- - Pacing pulse generation
- - Therapy pads and cables
- - Patient impedance
- - Capture versus electrical output
- - Pacing artifacts
- - Defibrillator analyzer testing
- - Common failure patterns
- - How to think through pacing troubleshooting
- The Simple Version
In transcutaneous pacing, the defibrillator sends repeated controlled-current pulses through the therapy cable and adhesive pads. The selected rate sets the timing, while the current setting and patient impedance affect the voltage required to deliver each pulse. Electrical output does not by itself prove that the myocardium responded; clinical capture is a separate patient response.
In demand mode, the defibrillator also monitors ECG activity. A detected intrinsic beat should inhibit the scheduled pacing pulse, while the absence of a qualifying beat allows a pulse to be delivered. ECG leads, sensing gain, artifact, cable condition, pads, load impedance, timing, and output must all be considered when pacing behavior appears wrong.
Worked Example: Pacing Pulses Are Missing
Use an approved defibrillator analyzer and the manufacturer's test mode and load. Confirm fixed versus demand mode, selected rate and current, therapy cable, and analyzer connection. If fixed pacing produces correct pulses but demand pacing does not, investigate ECG input, simulated rate, sensing threshold, lead selection, and inappropriate inhibition before blaming the pulse generator.
If neither mode produces output, inspect the complete therapy path and review error logs using the service procedure. Verify pulse rate, current across specified loads, demand inhibition, alarms, and cable condition after repair. High-energy circuits can retain hazardous charge; follow discharge and service-safety requirements.
What Is Transcutaneous Pacing?
Transcutaneous pacing uses external adhesive pads to deliver electrical pulses through the chest.
The goal is to stimulate the myocardium when the patient's own heart rate is inadequate.
It is different from:
- - Defibrillation
- - Synchronized cardioversion
even though the same therapy pads may be used.
Defibrillation vs Pacing
Defibrillation delivers:
One large, brief high-energy shock.
Pacing delivers:
Repeated lower-energy pulses.
The electrical output hardware and control method are different.
Pacing Rate
The pacing rate is usually expressed in:
Pulses per minute, often corresponding to beats per minute.
Example:
70 ppm.
The defibrillator should generate approximately:
- 70 pacing pulses each minute,
unless demand pacing suppresses some pulses because intrinsic cardiac activity is detected.
Pacing Current
Pacing output is commonly set in:
milliamps, or mA.
Example:
60 mA.
This controls the strength of the pacing pulse.
Current Is Different From Voltage
The clinician typically selects current.
The device adjusts the voltage needed to drive approximately that current through the patient or test load.
Why Voltage Changes
Ohm's law applies conceptually:
- Voltage = Current × Resistance
If patient impedance is higher:
The defibrillator may need more voltage to deliver the same pacing current.
Constant-Current Pacing
Many pacing systems are designed to approximate constant-current output.
That means:
Set:
60 mA.
The device attempts to deliver about:
- 60 mA
even as patient impedance varies within its supported range.
Current Source
Internally, the defibrillator uses controlled electronics to generate the pacing pulse.
This may involve:
- - Power conversion
- - Switching circuitry
- - Current regulation
- - Output monitoring
- Pacing Pulse
Each pacing pulse has several characteristics.
These may include:
- - Current
- - Duration
- - Repetition rate
- - Waveform shape
Manufacturer specifications define acceptable limits.
Pulse Width
Pulse width describes how long each pacing pulse lasts.
It may be measured in:
Milliseconds.
The pulse is much longer than many ordinary electronic control signals but far smaller than the interval between paced beats.
Why Pulse Width Matters
Stimulation depends on both:
- - Current
- - Pulse duration
A pulse with correct current but incorrect width may not represent correct pacing output.
Defibrillator Analyzer
A defibrillator analyzer can independently measure pacing output.
Depending on analyzer capability, it may measure:
- - Pacing current
- - Rate
- - Pulse width
- - Demand response
- - Refractory behavior
Set Current Is Not Proof of Delivered Current
Suppose the defibrillator is set:
60 mA.
That proves:
The commanded pacing current is 60 mA.
It does not prove:
60 mA is actually reaching the load.
Use an analyzer.
Example
Set:
60 mA.
Analyzer:
59 mA.
If within manufacturer tolerance:
Pass.
Low Output
Set:
60 mA.
Analyzer:
35 mA.
Now investigate:
- - Output circuitry
- - Therapy cable
- - Connector
- - Calibration
- - Load
- No Output
Pacing indicator flashes.
Analyzer measures:
0 mA.
Possible causes include:
- - Therapy cable
- - Output circuit
- - Pad detection
- - Internal switch
- Therapy Pads
Pacing current reaches the patient through adhesive therapy pads.
These may also support:
- - Defibrillation
- - Synchronized cardioversion
- - ECG monitoring
depending on system.
Therapy Cable
The cable is part of the pacing current path.
Damage can cause:
- - Reduced current
- - No current
- - Intermittent pacing
- - Pads-off message
- High-Current Path
A therapy cable may pass basic continuity and still have problems under actual output conditions.
Inspect and test using manufacturer-approved procedures.
Connector Damage
Look for:
- - Burn marks
- - Bent contacts
- - Cracks
- - Loose connection
Repeated high-energy therapy places demands on these interfaces.
Pad Contact
Poor pad contact increases impedance.
The defibrillator may:
- - Warn of poor contact
- - Limit output
- - Require higher voltage
- Patient Impedance
The electrical path through the chest presents impedance.
The device may monitor this to determine whether pacing output can be delivered effectively.
Analyzer Load
A defibrillator analyzer provides a known electrical load.
That gives a controlled way to verify pacing current.
Load Matters
If output is specified at a defined resistance:
Use that load.
A pacing circuit may behave differently at:
- - Low impedance
- - High impedance
- Multi-Load Testing
Some procedures may verify pacing current at more than one load.
That tests the current-regulation system.
Example
Set:
80 mA.
Analyzer at 50 Ω:
79 mA.
Analyzer at 100 Ω:
78 mA.
That suggests good regulation within those test conditions.
Current Drops at High Resistance
If:
- 50 Ω → 80 mA
- 100 Ω → 55 mA
the output stage may not be able to produce enough voltage under higher load.
Fixed Pacing
In fixed or asynchronous pacing, the device delivers pacing pulses at the selected rate regardless of intrinsic ECG activity.
Example
Set:
70 ppm.
The device sends approximately:
70 pulses per minute.
It does not wait for intrinsic beats.
Demand Pacing
Demand pacing monitors the patient's ECG.
If the patient's own heart beats faster than or appropriately within the pacing interval:
The device inhibits unnecessary pacing pulses.
Demand Mode Logic
Conceptually:
- Monitor ECG
- ↓
- Detect R wave
- ↓
- Reset pacing timer
- ↓
- Wait
If another R wave appears:
Reset again.
If no R wave appears before pacing interval expires:
Deliver pacing pulse.
ECG Sensing Is Critical
Demand pacing depends on reliable ECG detection.
If the ECG signal is poor, pacing timing can be wrong even if the pacing-output circuit is perfect.
False R-Wave Detection
Artifact may be mistaken for intrinsic cardiac activity.
The device may inhibit pacing unnecessarily.
Missed R Waves
If real R waves are not detected:
The device may pace even though intrinsic beats are present.
Demand Pacing Problem Can Be an ECG Problem
This is extremely important.
If fixed pacing works correctly but demand pacing behaves incorrectly:
Look at:
- - ECG leads
- - Signal quality
- - Lead selection
- - R-wave detection
before blaming the pacing-output circuit.
ECG Leads vs Therapy Pads
Some defibrillators may obtain ECG through:
- - Monitoring electrodes
- - Therapy pads
depending on configuration.
Know which source is being used for demand pacing.
Lead Selection
Demand pacing may require an ECG lead with clear R waves.
A poor lead selection can create unreliable sensing.
R-Wave Marker
Some devices display markers showing detected intrinsic beats.
Those markers can be useful clues.
If pacing is inhibited unexpectedly but the monitor shows false R-wave markers:
Sensing is the issue.
Pacing Artifact
Each pacing pulse produces a large electrical artifact on the ECG.
The monitor must distinguish:
- Pacing pulse
from:
Actual cardiac response.
Pace Marker
The display may show a marker for each pacing pulse.
This proves the device commanded or detected a pacing event.
It does not necessarily prove current reached the patient.
Electrical Capture
When a pacing pulse actually depolarizes the myocardium, this is called:
Electrical capture.
The ECG may show a QRS complex following the pacing spike.
Mechanical Capture
Mechanical capture means the electrical activation produces an actual cardiac contraction and pulse.
This is a clinical assessment.
Biomed Testing Does Not Prove Clinical Capture
A defibrillator analyzer can verify electrical pacing output.
It cannot prove that a specific patient's heart will respond to that current.
This Distinction Matters
If analyzer confirms:
- - Correct pacing current
- - Correct pulse width
- - Correct rate
then the equipment may be functioning correctly even if clinical capture was difficult.
Patient physiology is a separate factor.
Capture Threshold
Clinically, pacing current may be increased until capture occurs.
The required threshold varies by patient.
There is no single mA setting that guarantees capture in everyone.
Service Testing Is Different
During service:
Use defined analyzer loads and manufacturer specifications.
Do not judge device performance by whether a clinical patient happened to capture at a particular setting.
Pacing Rate Accuracy
Analyzer can verify pulse timing.
Example:
Set:
60 ppm.
Expected interval:
About 1 second between pulses.
Rate Error
Set:
60 ppm.
Analyzer:
45 ppm.
That is a timing/control problem.
Pulse Width Accuracy
Set pacing system may have a specified pulse width.
Analyzer verifies whether actual pulse duration falls within required range.
Why Rate and Current Should Be Tested Separately
A device can have:
- Correct current
but:
Wrong rate.
Or:
- Correct rate
but:
Low current.
Different functions.
Demand Pacing Test
An analyzer can generate a simulated ECG.
The defibrillator should detect the simulated beats and respond appropriately.
Demand Test Example
Defibrillator set:
70 ppm demand mode.
Analyzer provides ECG:
80 bpm.
Device should generally inhibit pacing because intrinsic simulated rate is faster.
Slow Simulated Rate
Analyzer provides:
40 bpm.
Defibrillator set:
70 ppm.
The device should pace as required between the slower simulated beats according to its timing logic.
Demand Timing
The exact timing behavior is manufacturer-specific.
Use the analyzer's pacing test and service manual requirements.
Refractory Period
After detecting a beat or delivering a pacing pulse, the device may temporarily ignore certain signals.
This helps prevent:
- - Double counting
- - Artifact sensing
- Noise Rejection
The pacing system must avoid interpreting:
- - Muscle artifact
- - Pacing artifact
- - Electrical noise
as true intrinsic beats.
ESU Interference
Electrical interference can affect ECG sensing.
The pacing-output system may be fine while demand sensing is disrupted.
Fixed Mode Can Help Isolate
If:
- Fixed pacing output is correct
but:
- Demand mode fails,
that strongly points toward:
- - ECG sensing
- - Timing logic
rather than the high-current output path.
No Pacing With Pads-Off Message
The device may intentionally prevent pacing if it does not detect a valid therapy connection.
Possible causes include:
- - Therapy pads
- - Cable
- - Impedance sensing
- - Connector
- Pad Detection
The system may test the electrical path before enabling therapy.
An open circuit may trigger:
Pads Off.
Defibrillation Works, Pacing Fails
Now the cable may still be good for the high-energy shock path while another pacing-specific circuit has failed.
Shared accessories do not always mean identical internal electronics.
Pacing Works, Defibrillation Fails
The reverse can also occur.
Pacing output does not prove the defibrillation high-voltage system works.
Pacing Output at Low Settings
Service procedures may test several current settings.
Example:
- 20 mA
- 100 mA
- 200 mA
depending on device.
Multi-Point Testing
One passing current does not prove the entire pacing-output range.
Linear Output Error
Example:
- Set → Analyzer
- 20 → 15
- 60 → 45
- 100 → 75
The system consistently delivers about 25% low.
That may suggest calibration or scaling.
Offset Error
Set → Analyzer
20 → 30
60 → 70
100 → 110
A roughly constant offset suggests a different problem.
High Current Only Fails
Example:
20 mA → Pass.
80 mA → Pass.
180 mA → Low.
The output stage may be reaching its voltage/current limit under load.
Battery Condition
Pacing may operate for extended periods on battery.
A weak battery could potentially affect overall therapy performance if the device cannot maintain internal supply rails.
Follow manufacturer battery and pacing-duration tests.
AC vs Battery Testing
If pacing works on AC but fails on battery:
Investigate:
- - Battery
- - Power path
before pacing hardware.
Continuous Load
Unlike a single defibrillation shock, pacing can continue for minutes or hours.
The output circuitry must handle repeated pulses reliably.
Thermal Behavior
A pacing fault may appear only after prolonged output.
If complaint is:
- Pacing stops after 20 minutes,
a brief analyzer test may miss it.
Long-Duration Test
Use the manufacturer-defined duration if required.
Observe:
- - Current stability
- - Errors
- - Temperature-related failures
- Internal Output Monitoring
The defibrillator may monitor its own pacing pulses.
If actual output differs from command, it may generate a fault.
Internal Self-Test
Self-test may check portions of:
- - Pacing output
- - Therapy circuit
But external analyzer verification remains important.
Internal Test Does Not Replace External Measurement
The device is checking itself using its own sensors.
Independent test equipment provides separate evidence.
Pacing Current Display
If the screen says:
- 80 mA,
that is usually the selected setting.
It is not necessarily a measured output value.
Therapy Event Log
The device may record:
- - Pacing mode
- - Rate
- - Current
- - Duration
This can be useful when investigating a clinical complaint.
Log Shows Setting, Not Necessarily Output
If the log says:
- Pacing 70 ppm / 80 mA,
that proves those settings were selected or recorded.
It does not independently prove the therapy cable actually delivered them.
Serious Clinical Events
If a defibrillator is involved in an event where pacing allegedly failed:
Preserve:
- - Logs
- - Accessories
- - Therapy cable
- - Device configuration
according to facility incident procedures.
Do not casually reset or update the device first.
Real-World Example: No Pacing Output
Device enters pacing mode.
Markers appear.
Analyzer:
0 mA.
Known-good therapy cable:
Normal current returns.
Failure follows cable.
Real-World Example: Low Current at High Load
Set:
100 mA.
50 Ω analyzer load:
99 mA.
100 Ω:
65 mA.
Output stage cannot maintain commanded current as required at higher impedance.
Real-World Example: Demand Pacing Erratic
Fixed mode:
Rate and current perfect.
Demand mode:
Pulses inhibited randomly.
ECG waveform shows heavy artifact and false R-wave markers.
Pacing hardware is fine.
Sensing is the problem.
Real-World Example: Paces Too Slowly
Set:
70 ppm.
Analyzer measures:
52 ppm.
Current and pulse width normal.
Timing/control system fails specification.
Real-World Example: “No Capture”
Clinical complaint:
Could not get capture even at high current.
Bench analyzer:
Pacing current, pulse width, and rate all within specification.
Equipment passes.
Clinical capture is patient-dependent and is not itself proof of device failure.
Real-World Example: Intermittent Therapy Connector
Pacing output drops when cable moves.
Defibrillation analyzer testing also shows intermittent therapy connection.
Connector damage found.
Common Mistakes
Treating Selected Current as Proof of Delivered Current
Measure with an analyzer.
Confusing Pacing Output With Clinical Capture
They are different questions.
Replacing the Pacing Circuit When Only Demand Mode Fails
Check ECG sensing first.
Assuming Pace Markers Prove Current Reached the Pads
They may only indicate a commanded pulse.
Ignoring Therapy Cable and Connectors
They are part of the current path.
Testing Only One Current Setting
Range problems may be missed.
Assuming Defibrillation Passing Proves Pacing Passing
They use different output modes.
Assuming Pacing Passing Proves Defibrillation Passing
Same problem in reverse.
A Useful Troubleshooting Framework
For a pacing complaint, ask:
- Does the device enter pacing mode?
Then:
- Is the therapy cable/pad path recognized?
Then:
- What current is selected?
Then:
- What current does an external analyzer measure?
Then:
- Is pacing rate correct?
Then:
- Is pulse width correct?
Then:
- Does fixed pacing work?
Then:
Does demand mode correctly detect and respond to simulated ECG?
That separates:
- - Output
- - Timing
- - Therapy path
- - ECG sensing
- Another Useful Question
Ask:
Is the device failing to generate the pacing pulse, failing to deliver it through the therapy path, or incorrectly deciding when to pace?
Those are three different problems.
What Did You Actually Prove?
If the screen shows:
- Pacing 70 ppm / 80 mA,
you proved:
Those pacing settings are selected.
If pacing markers appear:
You proved:
The device is generating or indicating pacing events internally.
You still have not proven:
80 mA is actually reaching the external load.
If a calibrated defibrillator analyzer measures:
- - Correct current
- - Correct rate
- - Correct pulse width
within manufacturer specification, you have strong evidence that the pacing-output system is functioning correctly under that test condition.
If demand testing also passes with simulated ECG:
You have additionally verified much of the pacing-sensing and timing path.
Final Thoughts for Biomeds
Defibrillator pacing is easiest to understand as two systems working together:
- ECG Sensing
and:
Electrical Pulse Delivery.
The delivery path is:
Pacing Circuit → Therapy Cable → Pads → Patient/Analyzer.
Demand pacing adds:
ECG → R-Wave Detection → Timing Logic → Pacing Decision.
So when pacing fails, first determine:
- Is the pulse missing?
- Is the pulse wrong?
Or is the machine deciding to pace at the wrong time?
And do not confuse:
- Electrical output
with:
Clinical capture.
A patient not capturing does not automatically prove the defibrillator failed.
Your analyzer tells you what the equipment actually delivered.
And as always:
- What did you actually prove?
- — Jake
- Important Note
Pacing waveform shape, current range, pulse width, demand algorithms, ECG sensing requirements, therapy accessories, impedance limits, and verification tolerances vary by defibrillator manufacturer and model. Follow current manufacturer service documentation, use approved therapy cables and calibrated defibrillator analyzers, and complete all required pacing, defibrillation, synchronization, alarm, battery, and safety testing before returning equipment to clinical use.
1/2
