3M Bair Hugger 775 Series

Error: FC-001 Internal Sensor 1 Over-Temperature / Hose Occlusion Fault

On this page

Asset Type

Patient Warming System

Manufacturer

3M

Model

Bair Hugger 775 Series

What This Guide Helps With

This guide helps troubleshoot Error FC-001 on the 3M Bair Hugger 775 Series Patient Warming System. FC-001 indicates an internal sensor 1 over-temperature condition and may also occur as a cautionary in-use fault when the hose end at the patient end is occluded.

This condition may interrupt patient warming and can be caused by hose occlusion, restricted airflow, incorrect hose or blanket setup, internal over-temperature sensing, or a sensor-related fault condition.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Patient warming should be transferred or stopped safely before troubleshooting begins.

Verify the Reported Error and Behavior

FC-001 may represent a true internal sensor over-temperature fault, but it can also occur from hose-end occlusion during use.

Disconnect Power and Allow the Unit to Reset

This allows the unit to power down fully and helps determine whether the condition was temporary.

Check for Hose Occlusion at the Patient End

A blocked hose can reduce airflow and cause localized heat buildup, which may trigger FC-001.

Inspect the Hose and External Air Path

Restricted airflow can create an over-temperature condition even if the heater and sensor system are functioning normally.

Check the Warming Blanket or Hose Card Setup

Improper setup can trap warm air and mimic a device fault.

Reconnect Power and Test Operation

If the hose occlusion was the cause, FC-001 should not return and airflow should remain normal.

Isolate User Setup vs Device Fault

Clinical Engineering should rule out external causes before assuming an internal device failure.

If the Problem Persists

If FC-001 remains active or returns after the hose occlusion is removed, external causes have been ruled out.

Persistent FC-001 may indicate an internal sensor over-temperature fault, sensor-related failure, airflow or heater control issue, or another internal device condition requiring service evaluation.

Clinical Use Tip

Never troubleshoot an over-temperature fault while the unit is actively warming a patient. Move the patient to a backup warming device first if warming is still clinically needed. Never rely on a patient warming unit that has an unresolved or repeating FC-001 fault.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bair Hugger 775 reported FC-001 during patient warming. Staff stated the fault occurred while the hose end was positioned near the patient blanket."

Cause

What was observed during troubleshooting.

Example:
"Found hose end occluded/mispositioned at the patient end, causing restricted airflow and cautionary FC-001 over-temperature condition. No permanent fault observed after correcting hose position."

Resolution

What action was taken.

Example:
"Disconnected unit from power, removed hose occlusion, repositioned hose correctly into hose card, waited 30–60 seconds, reconnected power, and tested operation. FC-001 did not return. Unit returned to service."

Helpful Details to Include (If Known)

Final Thought

FC-001 should be treated as a patient safety-related warming fault. Start with the simple and likely causes first: hose occlusion, hose position, airflow restriction, and setup. If the fault clears and does not return, document the correction clearly. If the fault persists, remove the unit from service and escalate for repair.

That is successful troubleshooting.