3M Bair Hugger 775 Series

Error: FC-006 Hose End Sensor 3 Shorted

On this page

Asset Type

Patient Warming System

Manufacturer

3M

Model

Bair Hugger 775 Series

What This Guide Helps With

This guide helps troubleshoot Error FC-006 on the 3M Bair Hugger 775 Series patient warming system. FC-006 indicates hose end sensor 3 is shorted and may be caused by a hose fault, damaged hose end sensor path, elbow sensor issue, moisture, contamination, or a device-level sensor circuit problem.

Because this fault involves temperature sensing, the unit should not be trusted for controlled patient warming until the problem is corrected and the warmer passes functional testing.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check the Hose Connection

Inspect the Hose for Damage or Contamination

Check for Hose End Sensor Failure

Check for Elbow Sensor Fault

Isolate Hose vs Device Fault

Verify Operation After Corrective Action

If the Problem Persists

If FC-006 remains after checking the hose connection, hose condition, patient-end hose sensor area, known-good hose substitution, and elbow sensor fault path, the common external causes have been ruled out.

A persistent FC-006 may indicate a device-level sensor circuit fault or internal failure condition.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot FC-006 on an active patient. Move the patient to a backup warming device first if warming therapy is needed. A patient warming system with an unresolved sensor fault should not be trusted for controlled heat delivery.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"3M Bair Hugger 775 Series displayed Error FC-006 Hose End Sensor 3 Shorted during use/startup and was removed from patient use."

Cause

What was observed during troubleshooting.

Example:
"Confirmed FC-006 fault. Inspected hose connection, hose condition, patient-end hose sensor area, and elbow sensor path. Fault appeared related to hose assembly, elbow sensor path, or internal sensor circuit depending on substitution testing."

Resolution

What action was taken.

Example:
"Removed unit from service, reseated and inspected hose, checked for damage or contamination, tested with known-good hose if available, replaced faulty hose or serviced elbow sensor path as indicated, completed calibration/functional verification, and returned to service only after FC-006 cleared."

Helpful Details to Include (If Known)

Final Thought

FC-006 should be treated as a real temperature-sensing fault until proven otherwise. Start with patient safety, verify the exact error, check the hose and simple external causes first, then isolate the hose, elbow sensor path, or device. If the fault persists after common checks, remove the warmer from service and escalate for repair. Good documentation protects the patient, the technician, and the next person troubleshooting the device.

That is successful troubleshooting.