3M Bair Hugger 775 Series

Error: FC-007 Hose End Sensor 2 Open / Hose Connection Fault

On this page

Asset Type

Patient Warming System

Manufacturer

3M

Model

Bair Hugger 775 Series

What This Guide Helps With

This guide helps troubleshoot Error FC-007 on the 3M Bair Hugger 775 Series Patient Warming System. FC-007 indicates hose end sensor 2 is open and may be caused by a disconnected hose, loose hose connection, faulty hose, damaged connector, elbow sensor issue, or related sensing circuit fault.

This error can affect hose temperature sensing and may prevent reliable warming operation. The unit should not be used for patient warming until the fault is understood, corrected, and verified.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Expected outcome: Patient warming therapy is maintained safely using reliable equipment.

Why it matters: A hose sensor fault may affect temperature monitoring or system control. Patient safety comes before equipment troubleshooting.

2. Verify the Reported Error and Behavior

Expected outcome: Error FC-007 is confirmed or can be reproduced.

Why it matters: This confirms a true system-reported hose sensor fault instead of a vague “not working” complaint.

3. Check Whether the Hose Is Disconnected

Expected outcome: The hose is securely connected and the error clears.

Why it matters: The troubleshooting note specifically points to the hose being disconnected from the unit. A loose hose can appear as an open sensor condition.

4. Inspect the Hose and Connector Area

Expected outcome: Hose and connection area are clean, intact, and properly seated.

Why it matters: FC-007 means hose end sensor 2 is open. A damaged hose, connector, or elbow sensor area can interrupt the sensor circuit.

5. Try a Known-Good Hose, If Available

Expected outcome: The error clears with a known-good hose.

Why it matters: If the fault clears with another hose, the original hose or hose sensor assembly is the likely cause.

6. Check for Elbow Sensor Issue

Expected outcome: Sensor-related fault clears after correcting the elbow sensor or hose sensor issue.

Why it matters: The system is reporting an open sensor condition, meaning the unit is not seeing a valid sensor circuit.

7. Calibrate and Verify Before Return to Service

Expected outcome: Unit passes calibration and functional verification.

Why it matters: Replacing sensing components without verifying operation can leave an unsafe or unreliable warming unit in service.

If the Problem Persists

If Error FC-007 remains after the hose is confirmed connected, the connector is inspected, a known-good hose is tested, and hose or elbow sensor issues are addressed, external and common causes have been ruled out.

At that point, the fault may involve the internal sensing circuit, connector interface, wiring path, or control electronics.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Do not keep cycling the unit back into use with an unresolved hose sensor fault.

Clinical Use Tip

Never troubleshoot a forced-air warming fault on an active patient. Move the patient to a backup warming device first if warming therapy is still needed, then continue troubleshooting the affected unit away from patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"3M Bair Hugger 775 displayed Error FC-007 during startup or use. Staff reported hose end sensor 2 open message and possible hose connection issue."

Cause

What was observed during troubleshooting.

Example:
"Hose connection was found loose at the warming unit, causing an open hose sensor condition. Error cleared after the hose was reseated."

Resolution

What action was taken.

Example:
"Reattached hose securely to warming unit. Powered unit on and verified Error FC-007 cleared. Ran functional check, confirmed normal operation, and returned unit to service."

Helpful Details to Include (If Known)

Final Thought

Error FC-007 should be approached as a hose sensing and connection problem first. Start with patient safety, verify the exact fault, check the hose connection, then isolate the hose and sensor path before suspecting deeper internal failure. Good documentation should clearly show what was checked, what corrected the issue, and whether the unit was safe to return to service.

That is successful troubleshooting.