Abbott Alinity c

ISE Module Calibration, Reference Solution, or Electrolyte Result Is Unstable

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Abbott

Model

Alinity c

What This Guide Helps With

Troubleshoot unstable electrolyte measurements or calibration failures involving ISE-related solutions, consumables, sample quality, and operating conditions.

Step-by-Step Troubleshooting

1. Protect Electrolyte Result Accuracy

Expected outcome: Affected electrolyte testing is isolated and clinical testing continuity is maintained.

2. Identify the Affected Electrolyte Measurement

Expected outcome: The scope and timing of the electrolyte measurement problem are established.

3. Verify ISE-Related Consumables

Expected outcome: Required ISE consumables are correct, available, and properly installed.

4. Review Recent Consumable Changes

Expected outcome: The consumable replacement history is verified and no required approved preparation step remains outstanding.

5. Evaluate Sample and QC Material Conditions

Expected outcome: Sample-related and control-material causes are ruled out or identified.

6. Check External Operating Conditions

Expected outcome: No unresolved external or environmental condition is interfering with ISE operation.

7. Perform Laboratory-Approved Calibration and QC

Expected outcome: Calibration is accepted and electrolyte QC is stable. Troubleshooting can stop when the laboratory approves the affected assays for patient use.

8. Escalate Unstable Measurements

Expected outcome: The unresolved electrolyte measurement issue is documented and escalated appropriately.

If the Problem Persists

After external consumables, QC materials, sample quality, and environmental conditions have been verified, remaining causes may involve internal ISE fluid handling, reference measurement components, electrodes, sensors, or calibration-related service faults.

The affected module should be:

Stopping when an ISE measurement remains unstable is proper troubleshooting.

Clinical Use Tip

Electrolyte results can influence urgent clinical treatment. Maintain access to another validated testing method until measurement stability is demonstrated.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported unstable sodium QC results and repeated ISE calibration difficulties."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering and laboratory personnel found that the installed ISE reference solution was expired."

Resolution

What action was taken.

Example:
"Laboratory personnel replaced the expired solution with an approved in-date product, completed the required preparation and calibration, and verified acceptable sodium QC."

Helpful Details to Include (If Known)

Final Thought

ISE instability requires careful review of solutions, sample conditions, and calibration history before internal components are suspected. Protect patient-result accuracy, avoid unsupported adjustments, verify corrective actions with QC, and escalate unresolved measurement problems.

That is successful troubleshooting.