Abbott Alinity c

Sample/Reagent Pipettor Collision, Clot, or Aspiration Error

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Abbott

Model

Alinity c

What This Guide Helps With

Troubleshoot pipettor collisions, clot detection, and aspiration failures caused by sample tubes, reagent cartridges, obstructions, or external handling conditions.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Stop Unsafe Operation

Expected outcome: Specimen processing is safely stopped or redirected, and the pipettor is not exposed to additional damage.

2. Confirm the Exact Pipettor Error

Expected outcome: The affected pipetting operation and failure pattern are identified.

3. Inspect Sample Tubes and Carrier Positioning

Expected outcome: The specimen is correctly positioned and suitable for the intended aspiration operation. If correcting a positioning problem restores verified operation, troubleshooting can stop.

4. Inspect Reagent Cartridges and Accessible Loading Areas

Expected outcome: No cartridge or accessible obstruction interferes with the pipettor path.

5. Check Pipettor Condition Without Manual Adjustment

Expected outcome: No visible pipettor damage is present, or suspected damage is identified for qualified service evaluation.

6. Verify Consumables and Operating Conditions

Expected outcome: External liquid-handling supplies and consumables are available, correctly installed, and free of reported faults.

7. Perform an Approved Functional Check

Expected outcome: The pipettor completes its operating sequence without errors and the laboratory accepts the verification results. Troubleshooting can stop.

8. Escalate Repeated Failures

Expected outcome: The unresolved fault is documented and transferred for manufacturer-level evaluation without additional mechanical damage.

If the Problem Persists

Once sample positioning, reagent seating, consumables, and visible obstructions have been ruled out, the problem may involve pipettor positioning, liquid-level sensing, aspiration components, motion control, or another internal liquid-handling mechanism.

The affected analyzer or module should be:

Knowing when to stop external troubleshooting is part of proper Clinical Engineering practice.

Clinical Use Tip

A pipettor collision can affect subsequent specimens. Coordinate with laboratory leadership to determine whether previously processed samples require review or repeat testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported repeated sample pipettor collision errors when processing a specific specimen rack."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a sample tube improperly seated in its carrier, allowing it to protrude into the pipettor travel path."

Resolution

What action was taken.

Example:
"The tube was correctly seated by laboratory personnel, the approved recovery sequence completed successfully, and subsequent functional checks and QC passed without additional errors."

Helpful Details to Include (If Known)

Final Thought

Pipettor troubleshooting begins with patient testing continuity and careful inspection of specimens, consumables, and positioning. Confirm external causes before suspecting internal mechanisms, stop when damage is possible, escalate appropriately, and document both the correction and final verification.

That is successful troubleshooting.