Abbott Precision Xceed Pro

Display or Keypad Malfunction

On this page

Asset Type

Blood Glucose Meter

Manufacturer

Abbott

Model

Precision Xceed Pro

What This Guide Helps With

Unresponsive buttons, frozen screen, dim display, or incorrect input preventing normal operation or patient testing.

Step-by-Step Troubleshooting

Ensure Patient Safety

Expected: No interruption to patient care
Why it matters: Prevents delays in glucose testing and incorrect results

Verify the Reported Issue

Expected: Issue is reproducible
Why it matters: Confirms a true malfunction vs. user error

Check Battery Status

Expected: Device powers on with stable display
Why it matters: Low or unstable power commonly causes display and keypad issues

Perform a Full Power Reset

Expected: Device reboots normally and responds to input
Why it matters: Clears temporary system freezes or software glitches

Inspect Keypad for Physical Obstruction or Damage

Expected: Buttons move freely and return properly
Why it matters: Contamination or wear can prevent proper electrical contact

Clean External Surfaces

Expected: Improved responsiveness if contamination was present
Why it matters: Residue buildup can interfere with button function

Check Display Visibility

Expected: Display is clear and readable
Why it matters: Physical screen failure will not be resolved with basic troubleshooting

Test with Known Good Dock or Accessories (if applicable)

Expected: Device operates normally outside faulty accessories
Why it matters: External accessories can sometimes interfere with operation

Attempt Basic Function Navigation

Expected: Device responds correctly to inputs
Why it matters: Confirms whether the issue affects all functions or only specific inputs

If the Problem Persists

All common external causes have been ruled out. The issue is likely internal (keypad membrane failure, display failure, or internal board issue).

Remove the device from service.
Label it Out of Service.
Send for repair or replace per facility protocol.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never attempt to “work around” unresponsive buttons during patient testing. Always switch to a functioning meter to avoid incorrect entries or missed results.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Device display frozen and keypad not responding during glucose testing."

Cause

What was observed during troubleshooting.

Example:
"Internal keypad/display failure after external checks and battery replacement."

Resolution

What action was taken.

Example:
"Removed from service and sent for repair; replacement meter issued to unit."

Helpful Details to Include (If Known)

Final Thought

Display and keypad issues often appear complex but are frequently caused by simple power or contamination problems. Always start with external checks and prioritize patient safety. When functionality is inconsistent or unresponsive, escalation is the correct decision. Proper documentation ensures clear communication and supports repair decisions.

That is successful troubleshooting.