Abbott Precision Xceed Pro

QC Lockout or Control Failures

On this page

Asset Type

Blood Glucose Meter

Manufacturer

Abbott

Model

Precision Xceed Pro

What This Guide Helps With

QC lockout, failed control tests, or meter preventing patient testing due to quality control requirements or invalid QC results.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient testing continues safely on alternate device
Why it matters: QC lockout is a safety feature and should not be bypassed

Verify the Reported Lockout or Failure

Expected: Clear indication of QC-related restriction
Why it matters: Confirms the issue is QC-related and not a device malfunction

Check QC Requirements and Timing Settings

Expected: QC requirement aligns with lockout condition
Why it matters: Many lockouts are triggered by normal policy settings, not failures

Inspect Control Solution (Most Common Cause)

Expected: Control solution is valid and within use period
Why it matters: Expired or degraded control solution commonly causes QC failures

Check Test Strips

Expected: Strips are valid and compatible
Why it matters: Strip issues are a leading cause of QC failure

Perform QC Test Correctly

Expected: QC result falls within acceptable range
Why it matters: Technique errors are a frequent cause of failed QC

Compare QC Results to Acceptable Range

Expected: Result is within defined QC limits
Why it matters: Out-of-range results trigger lockout by design

Try a New Bottle of Control Solution and New Strip Vial

Expected: QC passes with fresh materials
Why it matters: Eliminates degraded consumables as the cause

Check Meter Cleanliness and Strip Port Condition

Expected: Clean, unobstructed strip port
Why it matters: Contamination can affect strip readings and QC accuracy

Power Cycle the Meter

Expected: Meter resets and allows QC attempt
Why it matters: Clears minor software or temporary lock conditions

If the Problem Persists

If QC continues to fail after verifying control solution, strips, and technique, the issue is likely internal (measurement circuitry or calibration fault).

Remove the meter from service, label it Out of Service, and send for repair or bench evaluation. Do not attempt further troubleshooting beyond this point.

Clinical Use Tip

Never perform QC or troubleshooting while the device is in active patient use. Always switch to a verified meter before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Meter in ED locked out due to repeated QC failures and unable to perform patient testing."

Cause

What was observed during troubleshooting.

Example:
"Expired control solution and degraded test strip vial causing out-of-range QC results."

Resolution

What action was taken.

Example:
"Replaced control solution and test strips, performed QC successfully, and returned meter to service."

Helpful Details to Include (If Known)

Final Thought

QC lockouts are designed to protect patient safety, not hinder workflow. Most issues stem from consumables or technique, not device failure. Logical troubleshooting prevents unnecessary repairs and ensures reliable testing. Proper documentation supports both compliance and long-term device performance tracking.

That is successful troubleshooting.