Abbott Precision Xceed Pro

Strip Not Recognized or Insertion Errors

On this page

Asset Type

Blood Glucose Meter

Manufacturer

Abbott

Model

Precision Xceed Pro

What This Guide Helps With

Troubleshooting when test strips are not detected, improperly inserted, or causing repeated read errors due to external factors.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient testing continues safely on another device

Why it matters: Delayed or inaccurate glucose readings can impact clinical decisions

Verify Correct Strip Type

Expected: Correct strip type is used

Why it matters: Incorrect strips will not be recognized by the meter

Inspect Strip Orientation and Insertion Technique

Expected: Device detects strip and initiates test mode

Why it matters: Partial or reversed insertion prevents proper contact

Check Strip Condition and Expiration

Expected: New, clean, in-date strips function normally

Why it matters: Damaged or expired strips may not be recognized

Inspect the Strip Port for Contamination

Expected: Clean port allows proper electrical contact

Why it matters: Contamination blocks communication between strip and meter

Power Cycle the Device

Expected: Device resets and recognizes strip

Why it matters: Clears minor software or detection faults

Test with a New Strip from a Different Lot (If Available)

Expected: Device recognizes strip normally

Why it matters: Confirms whether issue is strip-related vs device-related

Observe for Repeated Failures or Intermittent Detection

Expected:

Why it matters: Helps differentiate accessory vs device failure

If the Problem Persists

If all external causes have been ruled out and strips are still not recognized, the issue is likely internal (e.g., strip port contact failure or internal circuitry issue).

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot strip recognition issues while relying on the device for active patient testing. Always transition to a functioning meter first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Meter not recognizing test strips during patient glucose testing."

Cause

What was observed during troubleshooting.

Example:
"Contamination found in strip port preventing proper strip contact."

Resolution

What action was taken.

Example:
"Cleaned strip port and verified proper operation with new test strip."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with simple, external checks like strip type, condition, and port cleanliness. This approach protects patient safety, avoids unnecessary repairs, and ensures accurate documentation when escalation is required.

That is successful troubleshooting.