Advanced Sterilization Products (ASP) AEROFLEX

Cycle Record, Barcode, Printer, or Tracking-System Export Fails

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Advanced Sterilization Products (ASP)

Model

AEROFLEX

What This Guide Helps With

Missing cycle documentation caused by barcode devices, printer supplies, cables, connectivity, configuration, tracking interfaces, or external infrastructure.

Step-by-Step Troubleshooting

1. Protect Traceability and Clinical Workflow

Do not rely on undocumented reprocessing when required cycle traceability cannot be established. Follow facility downtime documentation procedures while the electronic record pathway is unavailable.

Expected outcome: Endoscope traceability is maintained through an approved alternate documentation method.

2. Identify Which Documentation Function Failed

Determine whether the problem affects barcode scanning, local printing, stored cycle records, network export, or an external tracking system.

Expected outcome: The failed part of the documentation chain is isolated.

3. Confirm the Cycle Itself Completed Normally

Verify whether the underlying reprocessing cycle completed successfully and whether the issue is limited to documentation or export.

Expected outcome: A cycle-processing problem is not mistaken for a record-transfer problem.

4. Inspect Barcode Scanner and External Accessories

Check the scanner cable, connector, power state, scanning window, and physical condition. Test a known-readable barcode if available.

Expected outcome: The scanner reads and transfers data normally. If a loose connection or dirty scanner window was responsible, troubleshooting can stop after verification.

5. Check Printer Supplies and Connections

If printing is affected, verify approved paper or media is installed correctly, the printer is not jammed, covers are closed, and accessible cables or connections are secure.

Expected outcome: The printer accepts media and produces a complete readable record. If corrected, troubleshooting can stop.

6. Review Record Availability at the AER

Determine whether the completed cycle record is present locally even when it did not print or export. Do not delete or overwrite useful records during troubleshooting.

Expected outcome: The failure is isolated to output or transmission rather than cycle-record creation.

7. Inspect External Network or Tracking Connections

Check accessible Ethernet, USB, serial, or other approved external connections used by the facility. Look for disconnected, damaged, or loose cables and verify applicable network indicators.

Expected outcome: Physical connectivity is intact. If reconnecting an approved external cable restores export, troubleshooting can stop after a test record.

8. Compare Local Versus Infrastructure Failure

If possible, determine whether other AERs or connected devices can communicate with the same tracking system or network service.

Expected outcome: A facility-wide tracking or network outage is separated from a single AEROFLEX fault.

9. Perform a Controlled Documentation Test

Generate or use an approved test workflow to confirm barcode entry, local record creation, printing, and export as applicable.

Expected outcome: Required cycle documentation is correctly recorded and delivered to its intended destination.

10. Escalate Persistent Documentation Failure

If local records, peripherals, cables, and external infrastructure appear normal but required output still fails, stop external troubleshooting.

Expected outcome: The system is escalated to ASP service, IT, interface support, or the tracking-system vendor as appropriate.

If the Problem Persists

External causes involving scanner hardware, printer media, cables, local records, and facility connectivity have been ruled out. Remaining causes may involve software, configuration, the tracking interface, network services, or internal communication hardware.

Remove the AEROFLEX from service if required documentation cannot be maintained safely through an approved downtime process. Label it Out of Service when appropriate and arrange qualified repair or technical evaluation using ASP documentation and approved test equipment.

Return the system to normal workflow only after required traceability functions have been verified.

Clinical Use Tip

When electronic tracking is unavailable, use the facility-approved downtime documentation process so each scope remains traceable to its completed cycle.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported completed AEROFLEX cycles were recorded locally but were not exporting to the tracking system."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external network cable partially disconnected at the reprocessor."

Resolution

What action was taken.

Example:
"Reseated the cable, verified normal link indication, and confirmed a subsequent cycle record exported successfully."

Helpful Details to Include (If Known)

Final Thought

Traceability problems should be isolated one link at a time from record creation through final export. Verify external accessories and infrastructure before assuming an internal AER fault.

That is successful troubleshooting.