On this page
Asset Type
Manufacturer
Advanced Sterilization Products (ASP)
Model
What This Guide Helps With
Troubleshoot temperature-related processing problems caused by load conditioning, room environment, recent instrument processing, placement, chamber condition, or temperature-control faults.
Step-by-Step Troubleshooting
1. Protect Sterility Assurance
Do not release a load from a canceled or invalidated cycle associated with an unacceptable temperature condition.
Move time-sensitive instrument processing to another validated system.
Expected outcome: Patient instruments are not released from a cycle with questionable processing conditions.
2. Record the Exact Temperature-Related Condition
Document:
- Complete displayed message
- Whether the issue concerns the chamber or load
- When in the cycle it occurs
- Instruments involved
- Whether the condition repeats with every load
Do not assume a failed temperature sensor without first evaluating load and environment.
Expected outcome: The problem is clearly characterized.
3. Evaluate Recent Load History
Determine whether instruments were recently:
- Washed
- Heated
- Removed from another processing system
- Stored in a cold area
- Transported between substantially different environments
Coordinate corrective handling with Sterile Processing.
Expected outcome: The load has been appropriately conditioned before processing.
4. Inspect for Moisture and Condensation
Check the load for:
- Visible water
- Condensation
- Damp packaging
- Moisture retained in channels or complex devices
Temperature and moisture problems may occur together.
Expected outcome: The load is visibly dry.
5. Verify Load Arrangement
Confirm trays and packages are positioned correctly and are not crowded, stacked improperly, or contacting the chamber in an unusual manner.
Expected outcome: The load is appropriately arranged.
6. Check the Chamber Condition
With the sterilizer idle and safe, inspect the accessible chamber for:
- Moisture
- Debris
- Unusual residue
- Foreign material
Expected outcome: The chamber is clean and dry.
7. Check the Room Environment
Determine whether HVAC failure, room-temperature changes, sterilizer relocation, or prolonged shutdown could have affected the equipment or loads.
Do not adjust manufacturer calibration or temperature settings as an external troubleshooting step.
Expected outcome: No facility environmental condition explains the problem.
8. Compare With a Properly Conditioned Known-Good Load
When permitted, process an approved test or controlled load that is dry and at normal room conditions.
Expected outcome: The controlled load processes successfully. If so, the original load condition was responsible and equipment troubleshooting can stop.
9. Verify Successful Processing
After correcting the external condition, complete the required cycle and quality verification.
Expected outcome: The cycle completes without temperature-related cancellation and all required records and indicators are acceptable.
10. Escalate Repeated Temperature Faults
If correctly prepared and conditioned loads continue to trigger temperature faults, remove the sterilizer from service.
Expected outcome: Qualified service evaluates temperature sensing, chamber conditioning, controls, or related systems.
If the Problem Persists
Load conditioning, moisture, arrangement, chamber condition, and room environment have been ruled out. Remaining causes may involve temperature sensors, chamber-control functions, electrical components, software, or another service-level condition.
The sterilizer should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or service evaluation
- Evaluated using appropriate ASP documentation and approved test equipment
- Calibrated, repaired, or configured only by qualified personnel
Complete manufacturer-appropriate post-service verification before return to use.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Allow unusual hot or cold loads to reach appropriate processing conditions rather than repeatedly restarting canceled sterilization cycles.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported repeated STERRAD NX load-temperature cancellations with instruments brought directly from another processing area."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the problem followed recently heated loads while a properly conditioned control load processed normally."
Resolution
What action was taken.
Example:
"Coordinated appropriate load conditioning with Sterile Processing, verified a successful cycle with a properly prepared load, and returned the sterilizer to routine service."
Helpful Details to Include (If Known)
- Exact temperature-related message
- Chamber or load indication
- Instrument source before processing
- Load temperature history
- Visible moisture
- Load arrangement
- Chamber condition
- Room environmental changes
- Controlled-load result
- Final cycle result
- Final device status
Final Thought
Temperature-related cancellations should be investigated from the load and environment inward. Verify preparation and conditioning before assuming a sterilizer sensing or control failure.
That is successful troubleshooting.