On this page
Asset Type
Manufacturer
Advanced Sterilization Products (ASP)
Model
What This Guide Helps With
Troubleshoot repeat cycle cancellations during vacuum, sterilant delivery, diffusion, or plasma phases caused by load, packaging, consumables, chamber, or utility conditions.
Step-by-Step Troubleshooting
1. Protect Patient Care and Sterility Assurance
Treat every canceled sterilization cycle as unsuccessful unless the system and facility procedures explicitly establish otherwise. Do not release the load for patient use.
Route instruments through another validated sterilization process or verified sterilizer.
Expected outcome: Instruments from the canceled cycle remain segregated and patient care continues through an approved alternative.
2. Capture the Exact Cancellation Point
Record:
- Displayed message
- Phase when cancellation occurred
- Whether the cancellation happens at the same phase repeatedly
- Load contents
- Cycle selected
- Whether the unit previously completed cycles normally
Do not clear useful information before documenting it.
Expected outcome: The cancellation is characterized well enough to determine whether it follows the load, consumables, or sterilizer.
3. Inspect the Load and Packaging
Check for externally verifiable load problems such as:
- Excessive load density
- Items positioned against the chamber
- Incorrectly arranged trays
- Damaged packaging
- Restricted openings or channels
- Visible moisture
- Materials or packaging not approved for the selected process
Clinical Engineering should coordinate questionable load preparation with Sterile Processing rather than altering validated processing practices.
Expected outcome: The load is dry, correctly prepared, appropriately positioned, and compatible with the intended process.
4. Verify the Chamber and Door Area
With the system safe and idle, inspect accessible areas for:
- Debris
- Packaging trapped near the door
- Visible contamination
- Obstruction around the chamber opening
- Obvious seal damage visible without disassembly
Expected outcome: The chamber opening and accessible sealing surfaces are clean and unobstructed.
5. Verify Sterilant Cassette Condition
Confirm that an approved cassette is recognized, properly installed, undamaged, and shows usable capacity.
Do not attempt to reuse, refill, repair, or modify a questionable cassette.
Expected outcome: The sterilant source is recognized and available for the requested cycle.
6. Compare With a Controlled Load
If facility procedures allow, evaluate the sterilizer using a properly prepared controlled or routine load known to process successfully.
This helps distinguish a load-related cancellation from a sterilizer-related problem.
Expected outcome: A controlled load proceeds normally. If the original load configuration caused the cancellation, correct the processing issue and stop troubleshooting after successful verification.
7. Check Environmental and Utility Conditions
Verify that the sterilizer has stable electrical power and that required facility utilities and environmental conditions have not been interrupted or altered.
Check for recent:
- Power disturbances
- Facility work
- Utility interruptions
- Room environmental changes
- Relocation or service activity
Do not alter facility utilities or manufacturer configurations without authorization.
Expected outcome: No external facility condition explains the phase cancellation.
8. Observe for Abnormal External Behavior
During an approved test cycle, observe externally for:
- Unusual pump noise
- Repeated pauses
- Unexpected restart
- Door or chamber abnormalities
- Recurrent cancellation at the same phase
Do not bypass interlocks or operate with covers removed.
Expected outcome: The sterilizer progresses normally without abnormal external behavior.
9. Perform Final Cycle Verification
If the suspected external cause is corrected, complete the facility-required test or verification process before releasing the sterilizer for routine use.
Expected outcome: A complete cycle finishes normally with the expected cycle record and required quality indicators. Troubleshooting can stop.
10. Escalate Repeat Phase Cancellations
If cancellations recur with correctly prepared loads, recognized consumables, stable utilities, and no accessible obstruction, remove the sterilizer from service.
Expected outcome: Further investigation is transferred to qualified service personnel.
If the Problem Persists
External load, packaging, cassette, chamber-access, power, and environmental causes have been ruled out. Remaining possibilities include the vacuum system, valves, sterilant-delivery system, pressure sensing, plasma-generation system, control system, or another service-level condition.
The sterilizer should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or qualified service evaluation
- Evaluated using appropriate ASP documentation and approved test equipment
- Repaired or configured only by qualified personnel
Complete required post-repair functional and sterilization-process verification before return to service.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Any load from a canceled cycle must remain considered nonsterile until it completes an approved sterilization process.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported that the STERRAD NX repeatedly canceled during processing before completing the sterilization cycle."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found visible moisture in the prepared load, while the sterilizer completed an approved verification cycle with a properly dry load."
Resolution
What action was taken.
Example:
"Removed the wet load from processing, coordinated load correction with Sterile Processing, verified a complete successful cycle, and returned the sterilizer to service."
Helpful Details to Include (If Known)
- Exact cancellation message
- Phase where cancellation occurred
- Cycle selected
- Load contents and arrangement
- Visible moisture
- Cassette status
- Chamber and door condition
- Recent power or utility events
- Controlled-load result
- Final cycle result
- Final device status
Final Thought
A canceled sterilization cycle requires both patient-safety control of the load and logical isolation of load, consumable, chamber, utility, and equipment causes before service-level escalation.
That is successful troubleshooting.