Advanced Sterilization Products (ASP) STERRAD NX

Vacuum, Injection, Diffusion, or Plasma Phase Cancels

On this page

Asset Type

Sterilizer

Manufacturer

Advanced Sterilization Products (ASP)

Model

STERRAD NX

What This Guide Helps With

Troubleshoot repeat cycle cancellations during vacuum, sterilant delivery, diffusion, or plasma phases caused by load, packaging, consumables, chamber, or utility conditions.

Step-by-Step Troubleshooting

1. Protect Patient Care and Sterility Assurance

Treat every canceled sterilization cycle as unsuccessful unless the system and facility procedures explicitly establish otherwise. Do not release the load for patient use.

Route instruments through another validated sterilization process or verified sterilizer.

Expected outcome: Instruments from the canceled cycle remain segregated and patient care continues through an approved alternative.

2. Capture the Exact Cancellation Point

Record:

Do not clear useful information before documenting it.

Expected outcome: The cancellation is characterized well enough to determine whether it follows the load, consumables, or sterilizer.

3. Inspect the Load and Packaging

Check for externally verifiable load problems such as:

Clinical Engineering should coordinate questionable load preparation with Sterile Processing rather than altering validated processing practices.

Expected outcome: The load is dry, correctly prepared, appropriately positioned, and compatible with the intended process.

4. Verify the Chamber and Door Area

With the system safe and idle, inspect accessible areas for:

Expected outcome: The chamber opening and accessible sealing surfaces are clean and unobstructed.

5. Verify Sterilant Cassette Condition

Confirm that an approved cassette is recognized, properly installed, undamaged, and shows usable capacity.

Do not attempt to reuse, refill, repair, or modify a questionable cassette.

Expected outcome: The sterilant source is recognized and available for the requested cycle.

6. Compare With a Controlled Load

If facility procedures allow, evaluate the sterilizer using a properly prepared controlled or routine load known to process successfully.

This helps distinguish a load-related cancellation from a sterilizer-related problem.

Expected outcome: A controlled load proceeds normally. If the original load configuration caused the cancellation, correct the processing issue and stop troubleshooting after successful verification.

7. Check Environmental and Utility Conditions

Verify that the sterilizer has stable electrical power and that required facility utilities and environmental conditions have not been interrupted or altered.

Check for recent:

Do not alter facility utilities or manufacturer configurations without authorization.

Expected outcome: No external facility condition explains the phase cancellation.

8. Observe for Abnormal External Behavior

During an approved test cycle, observe externally for:

Do not bypass interlocks or operate with covers removed.

Expected outcome: The sterilizer progresses normally without abnormal external behavior.

9. Perform Final Cycle Verification

If the suspected external cause is corrected, complete the facility-required test or verification process before releasing the sterilizer for routine use.

Expected outcome: A complete cycle finishes normally with the expected cycle record and required quality indicators. Troubleshooting can stop.

10. Escalate Repeat Phase Cancellations

If cancellations recur with correctly prepared loads, recognized consumables, stable utilities, and no accessible obstruction, remove the sterilizer from service.

Expected outcome: Further investigation is transferred to qualified service personnel.

If the Problem Persists

External load, packaging, cassette, chamber-access, power, and environmental causes have been ruled out. Remaining possibilities include the vacuum system, valves, sterilant-delivery system, pressure sensing, plasma-generation system, control system, or another service-level condition.

The sterilizer should be:

Complete required post-repair functional and sterilization-process verification before return to service.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Any load from a canceled cycle must remain considered nonsterile until it completes an approved sterilization process.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported that the STERRAD NX repeatedly canceled during processing before completing the sterilization cycle."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found visible moisture in the prepared load, while the sterilizer completed an approved verification cycle with a properly dry load."

Resolution

What action was taken.

Example:
"Removed the wet load from processing, coordinated load correction with Sterile Processing, verified a complete successful cycle, and returned the sterilizer to service."

Helpful Details to Include (If Known)

Final Thought

A canceled sterilization cycle requires both patient-safety control of the load and logical isolation of load, consumable, chamber, utility, and equipment causes before service-level escalation.

That is successful troubleshooting.