On this page
Asset Type
Manufacturer
Advanced Sterilization Products (ASP)
Model
What This Guide Helps With
Troubleshoot abnormal vacuum performance by checking load, chamber, door sealing, environment, power, external airflow, and repeatability before service-level evaluation.
Step-by-Step Troubleshooting
1. Protect Sterility Assurance
If vacuum performance is unreliable or cycles are canceling, stop routine instrument processing on the affected sterilizer and redirect loads to verified equipment.
Expected outcome: No load is released from a cycle with questionable vacuum performance.
2. Characterize the Vacuum Complaint
Determine whether the sterilizer:
- Sounds louder than normal
- Takes unusually long during vacuum phases
- Cancels during vacuum
- Operates intermittently
- Shows the problem only with certain loads
Record any displayed message without assuming the pump itself is defective.
Expected outcome: The vacuum symptom is documented objectively.
3. Inspect the Load
Check for:
- Overloading
- Incorrect placement
- Packaging against the chamber
- Moisture
- Items that may interfere with evacuation
- Loads that differ significantly from those previously processed successfully
Expected outcome: The load is correctly prepared and arranged.
4. Inspect Chamber and Door Areas
With the unit safe and idle, inspect accessible areas for:
- Debris
- Packaging fragments
- Visible contamination
- Material interfering with the door seal
- Obvious seal abnormalities
Expected outcome: The chamber opening is clean and unobstructed.
5. Verify Stable Electrical Power
Confirm that the sterilizer is connected to its intended power source and that there is no evidence of intermittent power, loose external connection, or recent electrical disturbance.
Expected outcome: The sterilizer has stable facility power.
6. Check Accessible Ventilation Areas
Inspect external ventilation openings for obvious blockage, accumulated material, or objects restricting airflow.
Do not remove panels or access internal pump components.
Expected outcome: External ventilation is unobstructed.
7. Compare Empty or Controlled Testing
When approved by facility procedures, compare vacuum behavior using an appropriate controlled verification condition rather than the suspect clinical load.
Listen for major changes in sound and observe whether the cancellation repeats.
Expected outcome: Normal behavior with a controlled test points toward load-related conditions. Persistent abnormal behavior points toward the sterilizer.
8. Stop for Significant Mechanical Symptoms
Immediately discontinue testing if there is:
- Grinding
- Severe vibration
- Smoke
- Burning odor
- Fluid leakage
- Excessive heat
- Repeated electrical reset
Expected outcome: Potential equipment damage or safety risk is not worsened through repeated operation.
9. Verify Operation After External Correction
If an obstruction, load issue, or external environmental cause is corrected, perform the required verification cycle.
Expected outcome: Vacuum phases proceed normally without abnormal noise or cancellation. Troubleshooting can stop.
10. Escalate Persistent Vacuum Problems
If abnormal noise, slow evacuation, or vacuum-phase cancellation continues with proper loads and no external obstruction, remove the sterilizer from service.
Expected outcome: Qualified service evaluates the vacuum system using manufacturer-approved procedures.
If the Problem Persists
Common external causes such as load condition, chamber obstruction, door-area contamination, power problems, and ventilation restriction have been ruled out. Remaining possibilities include the vacuum pump, valves, pressure sensing, internal leaks, tubing, control circuitry, or another service-level condition.
The sterilizer should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or qualified service evaluation
- Evaluated using appropriate ASP documentation and approved test equipment
- Repaired only by qualified personnel
Complete all required post-repair cycle and sterilization-process verification before return to service.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
A cycle that cannot establish the required vacuum condition must not be treated as a successful sterilization cycle.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported that the STERRAD NX became unusually loud during the vacuum phase and repeatedly canceled cycles."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering confirmed abnormal vacuum noise and cancellation with a controlled load after external obstructions and load conditions were ruled out."
Resolution
What action was taken.
Example:
"Removed the sterilizer from service, labeled it Out of Service, and referred it for qualified vacuum-system evaluation before return-to-service testing."
Helpful Details to Include (If Known)
- Exact cycle message
- Phase of failure
- Description of abnormal noise
- Load condition
- Chamber and door inspection
- External ventilation condition
- Power condition
- Controlled-test result
- Presence of odor, heat, or vibration
- Final device status
Final Thought
Abnormal vacuum behavior should be verified without assuming a failed pump. Rule out external causes first, but stop testing promptly when persistent mechanical symptoms point to service-level repair.
That is successful troubleshooting.