AirLife FlowMate

Flow Continues With the Control Valve Fully Closed

On this page

Asset Type

Oxygen Flowmeter

Manufacturer

AirLife

Model

FlowMate

What This Guide Helps With

Troubleshoot oxygen flow that does not stop when the control valve is closed, including accessories, knob position, leakage, and valve-related problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Remove the Flowmeter From Active Therapy

An oxygen flowmeter that cannot reliably shut off should not remain in clinical use.

Transfer the patient to another verified oxygen source and remove the suspect flowmeter from the treatment setup.

Expected outcome: Patient oxygen delivery continues safely without relying on the defective shutoff function.

2. Confirm That Flow Is Actually Continuing

With an appropriate test setup, close the control valve using normal force.

Determine whether:

Do not overtighten the knob in an attempt to force shutoff.

Expected outcome: Continued oxygen flow or leakage is clearly distinguished from brief residual pressure in attached tubing.

3. Remove Outlet Tubing and Accessories

Disconnect tubing, humidifiers, adapters, and other outlet accessories.

Close the control valve again and reassess the outlet.

Expected outcome: Gas flow stops completely, or continued flow is confirmed as originating from the flowmeter rather than the downstream setup.

If the apparent flow was caused only by trapped gas or an accessory condition and normal shutoff is verified, proceed to final functional testing.

4. Inspect the Control Knob

Check whether the knob:

Do not remove or internally adjust the valve mechanism unless authorized by manufacturer service procedures.

Expected outcome: The external control reaches its intended closed position without damage or obstruction.

5. Verify Flowmeter Orientation and Mounting

Confirm that the unit is correctly installed and not being mechanically twisted or stressed by the wall connection or attached equipment.

Expected outcome: The flowmeter is securely mounted and external mechanical stress is not affecting control operation.

6. Compare With a Known-Good Flowmeter

If needed, use a compatible known-good flowmeter on the same verified oxygen source.

Confirm that the known-good device shuts off normally.

Expected outcome: The oxygen source behaves normally and the shutoff problem is isolated to the suspect flowmeter.

7. Treat Persistent Flow as a Failed Safety Function

If oxygen continues from the outlet with all external accessories removed and the valve normally closed, stop troubleshooting.

Do not compensate by overtightening the knob or relying on removal from the wall outlet as routine flow control.

Expected outcome: The device is identified as unreliable and removed from clinical availability.

8. Perform Final Functional Verification Only If the Cause Was External

If an external obstruction or accessory-related condition was corrected, verify:

Expected outcome: The control valve repeatedly shuts oxygen flow off normally.

If shutoff is reliable and all required checks pass, troubleshooting can stop.

If the Problem Persists

Common external causes such as residual pressure, outlet accessories, knob obstruction, mounting stress, and source problems have been ruled out.

Persistent flow with the valve closed may represent an internal control-valve, seat, seal, contamination, or other service-level gas-path problem.

The device should be:

Return-to-service testing should include reliable shutoff, controllable flow, leakage verification, connection integrity, and applicable manufacturer performance checks.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A flowmeter that will not completely shut off has an unreliable control function and should not be left installed for routine patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that oxygen continued flowing from the AirLife FlowMate even when the control knob was fully closed."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering confirmed continued outlet flow with all external tubing removed, isolating the problem to the flowmeter control assembly."

Resolution

What action was taken.

Example:
"The flowmeter was removed from service, labeled Out of Service, and sent for qualified bench evaluation and repair before return-to-service testing."

Helpful Details to Include (If Known)

Final Thought

Reliable shutoff is a basic safety requirement. Rule out trapped gas and accessories first, but persistent oxygen flow with the valve closed is the point to stop external troubleshooting and escalate the device.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide