AirLife FlowMate

Flow Reading Changes When the Outlet Tubing Is Connected

On this page

Asset Type

Oxygen Flowmeter

Manufacturer

AirLife

Model

FlowMate

What This Guide Helps With

Troubleshoot flow changes caused by restricted tubing, accessories, humidifiers, kinks, back pressure, connection problems, or an unreliable flowmeter.

Step-by-Step Troubleshooting

1. Protect the Patient and Use an Alternate Oxygen Source

If connecting tubing causes unpredictable oxygen flow, transfer the patient to another verified flowmeter or approved oxygen source before testing.

Do not troubleshoot fluctuating oxygen delivery while a patient depends on the affected setup.

Expected outcome: Oxygen therapy continues through a verified alternative while Clinical Engineering evaluates the suspect equipment.

2. Reproduce the Reported Condition

With an appropriate test setup, observe the indicated flow:

Document whether the float rises, falls, fluctuates, or becomes unstable.

Expected outcome: The change associated with connecting the tubing is confirmed.

3. Inspect the Oxygen Tubing

Examine the tubing for:

Ensure tubing is not trapped under equipment or furniture.

Expected outcome: Tubing is open, intact, appropriately connected, and free from visible restriction.

If replacing damaged tubing restores stable flow, troubleshooting can stop after final verification.

4. Inspect Attached Accessories

Check humidifiers, adapters, cannulas, masks, nebulizer interfaces, and other connected accessories as applicable.

Look for:

Use only accessories appropriate for the clinical setup.

Expected outcome: The downstream oxygen path is correctly assembled and free from obvious restriction.

5. Substitute Known-Good Tubing

Connect a compatible known-good length of tubing to the same flowmeter.

If the original setup included additional accessories, add them back one at a time to identify when the flow indication changes.

Expected outcome: Flow remains stable with known-good tubing, or the specific accessory causing the change is isolated.

If the flow remains stable after replacing the defective accessory or tubing, troubleshooting can stop after verification.

6. Verify the Flowmeter Is Upright and Secure

Confirm the flowmeter remains correctly oriented when tubing is connected.

Heavy tubing, humidifiers, or adapters can pull the flowmeter away from its proper operating orientation.

Expected outcome: The flowmeter remains upright and mechanically stable with the complete external setup attached.

7. Compare With a Known-Good Flowmeter

If the tubing and accessories appear normal, test the same external setup with a compatible known-good flowmeter on a verified oxygen source.

Expected outcome: The comparison helps determine whether the behavior follows the tubing/accessory setup or remains isolated to the original flowmeter.

8. Perform Final Functional Verification

After correcting the cause, confirm:

Use approved flow-measurement equipment when formal performance verification is required.

Expected outcome: The indicated and delivered oxygen flow remains stable with the intended external setup connected.

If normal performance is verified, troubleshooting can stop.

9. Escalate Unresolved Flow Changes

If the reading continues to change unexpectedly with verified tubing and accessories, remove the flowmeter from service.

Possible service-level causes should be evaluated on the bench rather than through further clinical-area troubleshooting.

Expected outcome: An oxygen flowmeter with uncertain performance is kept out of patient service.

If the Problem Persists

Common external restrictions, tubing defects, accessory problems, positioning issues, and connection problems have been ruled out.

Possible remaining categories include an internal valve problem, abnormal sensitivity to downstream pressure, internal gas-path restriction, or calibration/service-level failure.

The device should be:

Return-to-service testing should confirm flow accuracy, stability under appropriate outlet conditions, shutoff, and connection integrity.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Always evaluate the complete oxygen-delivery path; a normal-looking flowmeter cannot compensate for kinked or obstructed downstream tubing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the indicated oxygen flow dropped whenever patient tubing was connected to the AirLife FlowMate."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the oxygen tubing was sharply kinked behind the equipment and restricting the downstream flow path."

Resolution

What action was taken.

Example:
"The damaged tubing was replaced, and stable flow with the tubing connected, normal adjustment, and complete shutoff were verified."

Helpful Details to Include (If Known)

Final Thought

Trace the entire oxygen path before blaming the flowmeter. External tubing and accessories frequently change system behavior, so isolate them logically, verify performance, and escalate if reliable oxygen delivery cannot be demonstrated.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide