On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoot flow changes caused by restricted tubing, accessories, humidifiers, kinks, back pressure, connection problems, or an unreliable flowmeter.
Step-by-Step Troubleshooting
1. Protect the Patient and Use an Alternate Oxygen Source
If connecting tubing causes unpredictable oxygen flow, transfer the patient to another verified flowmeter or approved oxygen source before testing.
Do not troubleshoot fluctuating oxygen delivery while a patient depends on the affected setup.
Expected outcome: Oxygen therapy continues through a verified alternative while Clinical Engineering evaluates the suspect equipment.
2. Reproduce the Reported Condition
With an appropriate test setup, observe the indicated flow:
- With nothing connected to the outlet
- After connecting the reported tubing
- After connecting the complete accessory setup, when safe and appropriate
Document whether the float rises, falls, fluctuates, or becomes unstable.
Expected outcome: The change associated with connecting the tubing is confirmed.
3. Inspect the Oxygen Tubing
Examine the tubing for:
- Kinks
- Crushing
- Tight bends
- Internal moisture
- Visible contamination
- Damaged connectors
- Incorrect tubing
- Occluded ends
Ensure tubing is not trapped under equipment or furniture.
Expected outcome: Tubing is open, intact, appropriately connected, and free from visible restriction.
If replacing damaged tubing restores stable flow, troubleshooting can stop after final verification.
4. Inspect Attached Accessories
Check humidifiers, adapters, cannulas, masks, nebulizer interfaces, and other connected accessories as applicable.
Look for:
- Blockage
- Incorrect assembly
- Water accumulation
- Closed clamps
- Damaged fittings
- Misconnected ports
Use only accessories appropriate for the clinical setup.
Expected outcome: The downstream oxygen path is correctly assembled and free from obvious restriction.
5. Substitute Known-Good Tubing
Connect a compatible known-good length of tubing to the same flowmeter.
If the original setup included additional accessories, add them back one at a time to identify when the flow indication changes.
Expected outcome: Flow remains stable with known-good tubing, or the specific accessory causing the change is isolated.
If the flow remains stable after replacing the defective accessory or tubing, troubleshooting can stop after verification.
6. Verify the Flowmeter Is Upright and Secure
Confirm the flowmeter remains correctly oriented when tubing is connected.
Heavy tubing, humidifiers, or adapters can pull the flowmeter away from its proper operating orientation.
Expected outcome: The flowmeter remains upright and mechanically stable with the complete external setup attached.
7. Compare With a Known-Good Flowmeter
If the tubing and accessories appear normal, test the same external setup with a compatible known-good flowmeter on a verified oxygen source.
Expected outcome: The comparison helps determine whether the behavior follows the tubing/accessory setup or remains isolated to the original flowmeter.
8. Perform Final Functional Verification
After correcting the cause, confirm:
- Stable float position
- Predictable flow adjustment
- No unexpected change when tubing is attached
- No excessive mechanical stress on the flowmeter
- Flow can be shut off normally
- Connections remain secure
Use approved flow-measurement equipment when formal performance verification is required.
Expected outcome: The indicated and delivered oxygen flow remains stable with the intended external setup connected.
If normal performance is verified, troubleshooting can stop.
9. Escalate Unresolved Flow Changes
If the reading continues to change unexpectedly with verified tubing and accessories, remove the flowmeter from service.
Possible service-level causes should be evaluated on the bench rather than through further clinical-area troubleshooting.
Expected outcome: An oxygen flowmeter with uncertain performance is kept out of patient service.
If the Problem Persists
Common external restrictions, tubing defects, accessory problems, positioning issues, and connection problems have been ruled out.
Possible remaining categories include an internal valve problem, abnormal sensitivity to downstream pressure, internal gas-path restriction, or calibration/service-level failure.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
Return-to-service testing should confirm flow accuracy, stability under appropriate outlet conditions, shutoff, and connection integrity.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Always evaluate the complete oxygen-delivery path; a normal-looking flowmeter cannot compensate for kinked or obstructed downstream tubing.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the indicated oxygen flow dropped whenever patient tubing was connected to the AirLife FlowMate."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the oxygen tubing was sharply kinked behind the equipment and restricting the downstream flow path."
Resolution
What action was taken.
Example:
"The damaged tubing was replaced, and stable flow with the tubing connected, normal adjustment, and complete shutoff were verified."
Helpful Details to Include (If Known)
- Flow behavior before tubing connection
- Flow behavior after tubing connection
- Tubing type and condition
- Kinks or moisture observed
- Humidifier or accessories attached
- Known-good tubing results
- Known-good flowmeter comparison
- Device orientation
- Final flow verification
- Final device status
Final Thought
Trace the entire oxygen path before blaming the flowmeter. External tubing and accessories frequently change system behavior, so isolate them logically, verify performance, and escalate if reliable oxygen delivery cannot be demonstrated.
That is successful troubleshooting.