AirLife FlowMate

Flowmeter Tube Is Cracked, Clouded, or Contaminated

On this page

Asset Type

Oxygen Flowmeter

Manufacturer

AirLife

Model

FlowMate

What This Guide Helps With

Evaluate visible tube damage or contamination that may impair flow indication, structural integrity, cleanliness, or safe oxygen delivery.

Step-by-Step Troubleshooting

1. Protect the Patient and Remove the Suspect Flowmeter From Use

Do not continue oxygen therapy through a flowmeter with a cracked, severely clouded, contaminated, or otherwise compromised flow tube.

Transfer the patient to another verified oxygen source.

Expected outcome: Patient care continues through reliable equipment and the suspect flowmeter is isolated from clinical use.

2. Confirm the Visual Condition

Inspect the tube under good lighting and determine whether the issue is:

Expected outcome: The type and extent of the visible defect are clearly identified.

3. Inspect for Impact or Structural Damage

Examine the flowmeter body, tube supports, knob, inlet, and outlet for signs that the device was dropped or struck.

If a crack or structural damage is visible, do not pressurize the flowmeter merely to determine whether it still works.

Expected outcome: Structural damage is identified without unnecessarily placing the device under gas pressure.

If the tube is cracked, stop troubleshooting and remove the device from service.

4. Determine Whether the Contamination Is External

Inspect whether the observed material is on the outside surface or appears to be inside the flow tube.

Clean only external surfaces using facility-approved and manufacturer-compatible practices.

Do not introduce liquids, swabs, solvents, or tools into the flow tube unless specifically permitted by approved service documentation.

Expected outcome: External residue is removed safely, or internal contamination is identified for escalation.

5. Verify the Scale and Float Can Be Read Clearly

After permitted external cleaning, confirm that the scale markings and float position are clearly visible from the normal viewing position.

A tube that is too clouded or discolored to read reliably should not be used even if gas flow is present.

Expected outcome: The flow indication can be read clearly and unambiguously.

If visibility remains inadequate, remove the device from service.

6. Check for Moisture or Contamination From Downstream Accessories

Inspect attached humidifiers, tubing, and adapters for evidence that fluid may have migrated toward the flowmeter.

Correct or replace the external source of moisture or contamination before installing another flowmeter.

Expected outcome: The likely external source of contamination is identified and corrected where applicable.

7. Perform Functional Testing Only When the Tube Is Structurally Intact

If the tube has no crack or structural defect and the only issue was removable external contamination, use an appropriate test setup to verify:

Expected outcome: The flowmeter operates normally and the indication remains clearly visible.

If all checks pass and no internal contamination exists, troubleshooting can stop.

8. Escalate Internal Contamination or Tube Damage

Remove the flowmeter from service if any of the following remain:

Do not perform unauthorized tube replacement, internal cleaning, or recalibration.

Expected outcome: A flowmeter with compromised structural integrity or indication is prevented from returning to service.

If the Problem Persists

External surface contamination and accessory-related causes have been evaluated.

Persistent clouding, internal residue, moisture, cracking, or float impairment may require component replacement, internal cleaning under approved procedures, calibration verification, or other service-level work.

The device should be:

Before return to service, confirm structural integrity, clear visibility, free float movement, accurate flow performance, reliable shutoff, and leak-free connections.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

If the float or scale cannot be read confidently, the flowmeter should not be relied upon for setting patient oxygen flow.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the AirLife FlowMate tube appeared cloudy and the float was difficult to read."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found dried external residue covering the flow tube while the tube itself remained structurally intact and clear internally."

Resolution

What action was taken.

Example:
"The exterior was cleaned using approved practices, and clear visibility, free float movement, normal flow adjustment, shutoff, and functional performance were verified."

Helpful Details to Include (If Known)

Final Thought

A clear, intact flow tube is necessary for reliable oxygen-flow indication. Distinguish removable external contamination from structural or internal problems, verify performance only when safe, and escalate any device whose integrity or readability remains uncertain.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide