AirLife FlowMate Oxygen Flowmeter

No or Low Flow

On this page

Asset Type

Oxygen Flowmeter

Manufacturer

AirLife

Model

FlowMate

What This Guide Helps With

This guide addresses situations where the AirLife FlowMate oxygen flowmeter produces no oxygen flow or reduced flow. It focuses on external, easily verifiable causes first—such as gas supply, connections, and user settings—before assuming internal flowmeter failure.

Step-by-Step Troubleshooting

Verify Oxygen Source

Check Supply Pressure

Inspect Connections

Check Flowmeter Valve / Knob

Inspect Flowmeter Chamber and Indicator

Test with Alternate Tubing or Outlet

Observe for Leaks

If the Problem Persists

External causes have been ruled out (source pressure adequate, tubing clear, valves functional).

The flowmeter may have an internal valve or flow restriction failure.

Knowing when to escalate is proper troubleshooting and ensures patient safety.

Clinical Use Tip

Never troubleshoot an oxygen flowmeter while the patient is on active oxygen.

Move the patient to another verified oxygen delivery device first.

Ensure patient safety while verifying functionality.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“User reports ‘Oxygen flow is low or not present’ on the FlowMate flowmeter.”

Cause

What was observed during troubleshooting.

Example:
“Observed low source pressure from wall outlet; tubing kinked at elbow connector; flowmeter float stuck intermittently.”

Resolution

What action was taken.

Example:
“Straightened tubing, verified source pressure, cleaned float chamber, confirmed proper flow. Device returned to service. If unresolved, removed and sent for repair with ‘Out of Service’ label.”

Helpful Details to Include

Final Thought

Patient safety and logical, stepwise troubleshooting are paramount. Checking external sources, connections, and user settings first avoids unnecessary internal repair attempts. Documentation using CCR ensures clear communication and compliance. Escalate promptly when internal failure is suspected.

That is successful troubleshooting.