On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting failed USB export, image capture, video recording, or saved-file retrieval issues on the Ambu aView 2 Advance.
Step-by-Step Troubleshooting
Ensure Patient Safety First
Confirm the monitor is not being actively used for a patient procedure before troubleshooting recording, export, or capture issues.
If the monitor is needed for active visualization, provide another functional monitor before removing it from service.
Expected outcome: Patient care is not interrupted by troubleshooting activity.
Why it matters: Image capture and export are secondary functions, but the monitor must remain reliable for live visualization during procedures.
Verify the Reported Failure
Ask clinical staff what failed:
- Still image capture
- Video recording
- USB export
- File playback
- Saved image retrieval
- USB device recognition
- Export completes but files are missing or unreadable
Attempt to reproduce the issue using a test scope or approved test setup, not during patient use.
Expected outcome: The specific failure mode is confirmed.
If the issue cannot be reproduced and saved files/export operate normally, document findings and return the monitor to service.
Confirm the Monitor Is Fully Powered and Stable
Check that the aView 2 Advance is operating on stable power.
If using AC power, verify the power adapter is fully seated and connected to a known-good outlet.
If operating on battery, confirm the battery has adequate charge before attempting recording or export.
Expected outcome: The monitor remains powered without interruption during capture, recording, or export.
Why it matters: File writing or USB export may fail if the monitor loses power or shuts down during the process.
Check Available Storage or Recording Capacity
Review whether the monitor has available internal storage or whether prior recordings/images are filling memory.
Delete old non-required test files only if allowed by facility policy and clinical documentation requirements.
Expected outcome: The monitor has sufficient storage available for new images or recordings.
If storage was full and clearing approved files restores capture or recording, document the resolution and stop.
Verify Correct Capture or Recording Workflow
Confirm the user is pressing the correct image capture or recording control and allowing the action to complete.
Watch for on-screen confirmation, recording indicators, file-save messages, or error messages.
Expected outcome: The monitor confirms that the image or video was saved.
If no save confirmation appears, continue troubleshooting.
Check the Connected Endoscope or Camera Source
Confirm that a compatible Ambu endoscope is connected and that a live image is displayed before attempting capture or recording.
Reseat the scope connector if safe to do so during testing.
Inspect the connector area for contamination, bent contacts, loose fit, or obvious damage.
Expected outcome: A stable live image is present before capture or recording is attempted.
Why it matters: Some capture or recording functions may fail or be unavailable if the monitor does not recognize a valid video source.
Power Cycle the Monitor
Turn the monitor off using the normal shutdown method.
Disconnect accessories and USB devices.
Wait briefly, then power the monitor back on and reconnect only the required test scope.
Attempt image capture and recording again.
Expected outcome: Temporary software or file-handling issues clear after restart.
If capture or recording works after restart, document the reset and verify repeated operation before returning to service.
Inspect the USB Port Externally
Check the USB port for:
- Bent or recessed contacts
- Broken plastic insert
- Debris or contamination
- Loose fit
- Signs of liquid exposure
- Evidence the USB device was forced in incorrectly
Do not insert tools into the port unless following approved cleaning procedures.
Expected outcome: The USB port is clean, intact, and physically secure.
If the port is damaged, remove the monitor from service and send it for repair evaluation.
Try a Known-Good USB Drive
Use a known-good USB flash drive that is approved by the facility.
Avoid encrypted, password-protected, damaged, or unusually large-capacity drives if they are not known to work with the device.
Insert the drive fully and wait for the monitor to recognize it.
Expected outcome: The monitor detects the USB drive and allows export.
If export works with a different USB drive, the original USB device is the likely cause. Document the finding and stop.
Check USB Drive Format and Available Space
Verify the USB drive has available storage space and is formatted in a compatible format per facility practice and manufacturer guidance.
Remove unrelated files if needed or use a clean approved USB drive.
Expected outcome: The USB device has enough capacity and can accept exported files.
If a full or incompatible USB drive was the issue, replace it with an approved drive and document the resolution.
Repeat Export With a Small Test File
Capture a new test image or short test recording.
Attempt to export only that file.
Expected outcome: A small test file exports successfully.
Why it matters: Large recordings may fail due to storage limits, file corruption, USB drive capacity, or export timeout conditions.
Check for File Visibility on a Computer
After export, safely remove the USB drive if the monitor provides that option.
Insert the USB drive into a hospital-approved computer and confirm whether files are present and readable.
Expected outcome: Exported files appear on the USB drive and open normally.
If files are present but cannot be opened, document the file type, file size, and computer/application used.
Check Date, Time, and File Naming Behavior
Verify the monitor date and time are correct.
Check whether files are saved under unexpected timestamps, folders, or naming formats.
Expected outcome: Captured images and recordings can be located using the correct date/time and file structure.
If incorrect date/time caused staff to think files were missing, correct the settings if permitted and document the finding.
Test With Another Known-Good Monitor or Accessory Setup
If available, compare operation using another Ambu aView 2 Advance monitor, the same USB drive, and the same test scope.
Expected outcome: The issue follows either the monitor, USB device, or accessory setup.
If the same USB drive works on another monitor but not this unit, suspect a monitor USB/export fault.
Review for Software Lockup or Repeated Error Messages
Note any recurring messages, frozen screens, failed save confirmations, export progress failures, or reboot behavior.
Do not attempt software reloads, internal service procedures, or board-level repair unless authorized by service documentation and facility policy.
Expected outcome: Any repeatable software or export fault is documented clearly for repair escalation.
If the Problem Persists
If capture, recording, or USB export still fails after power, storage, USB drive, file size, workflow, and accessory checks are completed, the most likely cause is an internal monitor fault, software/file system issue, damaged USB interface, or storage-related failure.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
Knowing when to stop is proper troubleshooting. Do not continue using a monitor for clinical documentation if captured images or recordings cannot be reliably saved or exported.
Clinical Use Tip
Do not troubleshoot recording or export problems during an active patient procedure. If live visualization is required, move the procedure to another compatible monitor first. Documentation features should be tested only when patient safety and workflow allow.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Ambu aView 2 Advance would not export captured procedure images to a USB drive."
Cause
What was observed during troubleshooting.
Example:
"Verified that the original USB drive was not recognized by the monitor, while a known-good approved USB drive exported test images successfully."
Resolution
What action was taken.
Example:
"Replaced the USB drive with an approved working device, confirmed image capture and export, and returned the monitor to service."
Helpful Details to Include (If Known)
- Whether the issue affected image capture, video recording, USB export, or file playback
- Exact error message or on-screen behavior
- Whether live endoscope image was present
- USB drive brand, size, and format if known
- Whether a known-good USB drive was tested
- Whether internal storage appeared full
- Whether old files were deleted per policy
- Whether files were visible on a computer after export
- Date/time settings verified
- USB port condition
- Any freezing, rebooting, or failed save/export messages
- Final device status
Final Thought
USB, recording, and image capture failures should be approached logically: confirm patient safety, verify the reported symptom, test power and storage, rule out the USB device, and document the exact failure. If external causes are ruled out, escalation is the safest and most appropriate Clinical Engineering response.
That is successful troubleshooting.