Arjo Flowtron Excel

Pressure alarm conditions

On this page

Asset Type

Sequential Compression Device (SCD)

Manufacturer

Arjo

Model

Flowtron Excel

What This Guide Helps With

Pressure alarms caused by blocked tubing, leaks, incorrect garments, or connection issues affecting proper inflation and therapy delivery.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm Condition

Check Sleeve (Garment) Connection

Inspect Sleeve and Tubing for Leaks or Damage

Check for Kinks or Obstructions in Tubing

Verify Correct Garment Type and Size

Test Alternate Sleeve (Known Good)

Inspect Pump Ports for Debris or Damage

Perform Power Cycle

If the Problem Persists

If all external causes have been ruled out, the issue is likely internal (pressure sensor, internal pump, or control board).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Do not troubleshoot pressure alarms while the device is connected to a patient. Always transfer the patient to a functioning SCD first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit alarming for pressure during use; sleeves not inflating properly."

Cause

What was observed during troubleshooting.

Example:
"Leaking sleeve tubing preventing adequate pressure buildup."

Resolution

What action was taken.

Example:
"Replaced sleeve with known-good set; verified normal inflation and no alarms."

Helpful Details to Include (If Known)

Final Thought

Pressure alarms are most commonly caused by external airflow issues. A methodical approach—starting with connections, tubing, and garments—ensures patient safety and avoids unnecessary repairs. Recognizing when the issue is internal is just as important as fixing simple problems.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide