Arjo Flowtron Excel

Uneven Inflation Between Chambers

On this page

Asset Type

Sequential Compression Device (SCD)

Manufacturer

Arjo

Model

Flowtron Excel

What This Guide Helps With

Uneven or inconsistent chamber inflation caused by tubing issues, garment defects, incorrect setup, or external pneumatic restrictions.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Check Garment (Sleeve) Application

Inspect Tubing Connections

Check Tubing for Kinks or Obstructions

Inspect Garment for Internal Defects

Swap with a Known-Good Garment

Check Pump Output Ports

Power Cycle the Unit

If the Problem Persists

If uneven inflation continues after verifying tubing, garment, and connections, the issue is likely internal (valve control, pressure regulation, or pneumatic distribution failure).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and reflects proper troubleshooting.

Clinical Use Tip

Do not attempt to troubleshoot while the device is in use on a patient. Always transition the patient to a functioning SCD before evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported SCD sleeve inflating unevenly between chambers during therapy."

Cause

What was observed during troubleshooting.

Example:
"Faulty garment with internal chamber restriction identified during testing."

Resolution

What action was taken.

Example:
"Replaced sleeve and verified normal sequential inflation across all chambers."

Helpful Details to Include (If Known)

Final Thought

Uneven chamber inflation is most often caused by external factors like garment condition or tubing restriction. Systematic troubleshooting ensures patient safety, prevents unnecessary repairs, and supports accurate documentation. Escalate only after external causes are ruled out.

That is successful troubleshooting.