Arjo Hydroven Series

Cycle Timing Errors

On this page

Asset Type

Sequential Compression Device (SCD)

Manufacturer

Arjo

Model

Hydroven Series

What This Guide Helps With

Abnormal inflation/deflation timing, inconsistent cycle duration, or therapy not following expected sequence due to setup, garment, or external airflow issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Confirm Correct Therapy Mode/Settings

If corrected and timing normalizes → stop troubleshooting.

Check Garment (Sleeve) Application

If corrected and timing normalizes → stop troubleshooting.

Inspect Tubing Connections

If corrected and timing normalizes → stop troubleshooting.

Check for Kinks or Restrictions in Tubing

If corrected and timing normalizes → stop troubleshooting.

Inspect Garment and Tubing for Leaks

If leak found and corrected → stop troubleshooting.

Swap with Known-Good Garment and Tubing

If resolved → faulty accessory identified.

Observe for Consistency Across Multiple Cycles

If the Problem Persists

If all external causes have been ruled out and cycle timing remains inconsistent, the issue is likely internal (control board, timing circuitry, or internal pneumatic regulation).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot while the device is in use on a patient. Always transition the patient to a functioning SCD before evaluating timing issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"SCD reported to have irregular compression timing and inconsistent cycle duration."

Cause

What was observed during troubleshooting.

Example:
"Faulty garment tubing causing intermittent airflow restriction affecting cycle timing."

Resolution

What action was taken.

Example:
"Replaced tubing and garment; verified normal cycle timing across multiple cycles."

Helpful Details to Include (If Known)

Final Thought

Cycle timing issues are often caused by airflow disruptions rather than internal failure. A structured approach that prioritizes simple external checks ensures patient safety and prevents unnecessary repairs. Clear documentation supports future troubleshooting and reliability tracking.

That is successful troubleshooting.