Arjo Hydroven Series

Garment Not Recognized

On this page

Asset Type

Sequential Compression Device (SCD)

Manufacturer

Arjo

Model

Hydroven Series

What This Guide Helps With

Garment not detected or therapy not starting due to connection issues, incompatible sleeves, or external pneumatic faults.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Check Garment Connection

Inspect Tubing and Connectors

Confirm Correct Garment Type

Swap with a Known Good Garment

Inspect for Air Leaks During Startup

Perform a Power Cycle

If the Problem Persists

All common external causes have been ruled out. The issue is likely internal (pressure sensing, port detection, or control board fault).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot garment recognition issues while connected to a patient. Always transition the patient to another device before testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"SCD not recognizing garment and therapy not starting."

Cause

What was observed during troubleshooting.

Example:
"Faulty garment tubing preventing proper pressure detection."

Resolution

What action was taken.

Example:
"Replaced garment with known good unit; device function restored."

Helpful Details to Include (If Known)

Final Thought

Start with simple connection and accessory checks before assuming device failure. Most garment recognition issues are external. Knowing when to escalate ensures patient safety and efficient workflow. Proper documentation supports future troubleshooting and trending.

That is successful troubleshooting.