Arjo Maxi Sky Series

Lift not moving along track

On this page

Asset Type

Patient Lift

Manufacturer

Arjo

Model

Maxi Sky Series

What This Guide Helps With

Lift not traveling along ceiling track due to power, control, obstruction, or track-related issues affecting horizontal movement.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Check Emergency Stop (E-Stop)

If lift moves → stop troubleshooting.

Verify Battery Charge and Seating

If movement returns → stop troubleshooting.

Test Hand Control Functions

If no functions work → troubleshoot hand control/power instead.

Inspect Track for Obstructions

If obstruction found and removed → retest and stop if resolved.

Check for Track Misalignment or Damage

If damage is found → remove from service.

Inspect Trolley Wheels and Movement

If binding is present → likely mechanical issue.

Check Track Charging/Power Rail (if applicable)

If dirty → clean and retest.

Cycle Power to the Unit

If movement returns → stop troubleshooting.

If the Problem Persists

Common external causes have been ruled out. The issue is likely internal (track motor, drive assembly, or control system failure).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and protects patient safety.

Clinical Use Tip

Never troubleshoot a ceiling lift while a patient is suspended. Always transfer the patient safely before evaluating movement issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Lift would not travel along ceiling track when using hand control."

Cause

What was observed during troubleshooting.

Example:
"No external obstruction found; suspected internal track drive motor failure."

Resolution

What action was taken.

Example:
"Removed lift from service and sent to vendor for repair evaluation."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with simple, visible checks before assuming internal failure. Ceiling lift movement issues are often caused by power or track conditions, but recognizing when escalation is needed is just as important. Proper documentation supports both safety and long-term reliability.

That is successful troubleshooting.