Arjo Maxi Sky Series

Track obstruction or trolley binding

On this page

Asset Type

Patient Lift

Manufacturer

Arjo

Model

Maxi Sky Series

What This Guide Helps With

Lift not moving smoothly along track due to obstruction, debris, misalignment, or trolley resistance affecting horizontal travel.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Check Emergency Stop (E-Stop)

If movement returns → stop troubleshooting.

Inspect Track for Obstructions

If obstruction is found and removed → retest. If resolved, stop.

Check for Physical Damage or Misalignment

If damage is found → remove from service and escalate.

Inspect Trolley Wheels and Movement

If abnormal noise or resistance is present → escalate.

Check for Load or Sling Interference

If interference is removed and movement returns → stop troubleshooting.

Verify Battery and Power Output

If movement improves → stop troubleshooting.

Check for Environmental Factors

If cleared and movement returns → stop troubleshooting.

If the Problem Persists

Common external causes have been ruled out. The issue is likely internal (trolley assembly, bearings, or drive system).

Remove the lift from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents further damage and ensures safety.

Clinical Use Tip

Do not troubleshoot while a patient is suspended. Always transfer the patient safely before evaluating lift movement issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Lift not moving along track and stopping intermittently during transport."

Cause

What was observed during troubleshooting.

Example:
"Debris found in ceiling track causing trolley binding."

Resolution

What action was taken.

Example:
"Removed obstruction, cleaned track, and verified smooth operation."

Helpful Details to Include (If Known)

Final Thought

Track-related issues are often caused by simple external factors like debris or obstruction. A structured approach ensures safe, efficient troubleshooting while avoiding unnecessary repairs. Always prioritize patient safety and proper escalation when mechanical resistance remains.

That is successful troubleshooting.