Arjo Sara Stedy

Brake Failure or Caster Locking Issues

On this page

Asset Type

Patient Lift

Manufacturer

Arjo

Model

Sara Stedy

What This Guide Helps With

Brakes not holding, caster not locking/unlocking, or lift moving unintentionally due to debris, wear, or mechanical issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Inspect for Debris Around Casters

Check Brake Pedal/Lever Function

Inspect Caster Condition

Verify Floor Surface Conditions

Check Mounting and Stability of Casters

Test All Casters Together

If the Problem Persists

If all external causes have been ruled out and brakes still fail to hold or function properly, the issue is likely due to internal mechanical wear within the caster assembly.

Remove the device from service, label it Out of Service, and send for repair or caster replacement.

Knowing when to stop prevents unsafe equipment from remaining in use.

Clinical Use Tip

Never troubleshoot brake or caster issues with a patient on the device. Always transfer the patient first to eliminate fall risk.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported lift rolling even with brakes engaged."

Cause

What was observed during troubleshooting.

Example:
"Right rear caster brake worn and not fully engaging wheel."

Resolution

What action was taken.

Example:
"Replaced faulty caster and verified proper brake function."

Helpful Details to Include (If Known)

Final Thought

Brake and caster issues directly impact patient safety and should never be minimized. Start with simple mechanical checks, rule out environmental factors, and escalate appropriately. Proper documentation ensures accountability and repeat issue tracking.

That is successful troubleshooting.