B. Braun Dialog+

Ultrafiltration Rate Incorrect or Fluid Balance Test Failure

On this page

Asset Type

Hemodialysis (HD) Machine

Manufacturer

B. Braun

Model

Dialog+

What This Guide Helps With

Troubleshoot ultrafiltration or fluid-balance problems caused by setup, tubing, dialysate flow, supply conditions, configuration, leakage, or internal hydraulic issues.

Step-by-Step Troubleshooting

1. Protect the Patient From Incorrect Fluid Removal

Do not continue dialysis if ultrafiltration delivery or fluid balance cannot be trusted. Follow clinical procedure and transfer the patient to another verified dialysis machine when necessary.

Keep the affected Dialog+ out of patient service until fluid-management performance is verified.

Expected outcome: The patient is protected from unintended fluid removal or inadequate ultrafiltration.

2. Confirm the Exact Complaint

Determine whether the reported problem is incorrect displayed ultrafiltration rate, unexpected patient fluid removal, inability to reach the programmed rate, fluid-balance test failure, or repeated balance-related alarms.

Document the programmed settings and observed behavior without changing the prescription merely to clear the problem.

Expected outcome: The fluid-management complaint is clearly defined.

3. Verify Treatment or Test Settings

Review accessible ultrafiltration-related settings and confirm they match the intended prescription or approved bench-test setup.

Look for data-entry errors before assuming a machine malfunction.

Expected outcome: The configured target and rate are correct. If correcting an entry error resolves the issue and all required checks pass, troubleshooting may stop.

4. Inspect External Dialysate and Drain Connections

Check treated-water, dialysate, concentrate, and drain hoses for kinks, restriction, leakage, loose fittings, improper routing, or connections that may affect hydraulic balance.

Expected outcome: External fluid connections are secure, unrestricted, and dry. If an external flow restriction caused the failure and correction restores normal testing, troubleshooting may stop.

5. Check for Visible Leaks

Inspect the machine exterior, accessible tubing, connectors, and surrounding floor for water or dialysate leakage.

If an internal leak is suspected or fluid is emerging from the cabinet, remove the machine from service rather than continuing operation.

Expected outcome: No external or suspected internal fluid leak is present.

6. Verify Water and Drain Infrastructure

Confirm treated-water supply and drain conditions are normal. Determine whether other machines on the same water or drain infrastructure are experiencing similar issues.

Expected outcome: The problem is isolated to the machine or an infrastructure issue is identified for appropriate escalation.

7. Repeat the Approved Fluid Balance Test

With no patient connected, repeat the applicable manufacturer-approved preparation or fluid-balance verification after external causes have been corrected.

Do not bypass a failed balance test.

Expected outcome: The machine successfully completes its fluid-balance check. If successful and no other faults remain, troubleshooting may stop after final verification.

8. Verify Ultrafiltration Performance With Approved Test Equipment

If within Clinical Engineering scope, use approved dialysis test equipment and manufacturer procedures to evaluate fluid-removal performance.

Do not alter calibration values merely to force a passing result.

Expected outcome: Measured performance meets applicable manufacturer requirements.

9. Escalate Persistent Fluid-Balance Failure

If settings, external fluid paths, treated-water supply, drain conditions, and visible leaks are ruled out but balance testing still fails, stop external troubleshooting.

Expected outcome: Internal hydraulic, balancing, valve, flow-measurement, calibration, or control problems are referred for qualified service.

If the Problem Persists

Common external causes have been ruled out. Remaining possibilities include internal hydraulic balancing components, flow measurement, valve control, calibration, leakage, or another service-level problem.

The device should be:

Complete all required fluid-balance, ultrafiltration, alarm, and functional tests before returning the machine to service.

A failed fluid-balance test should not be bypassed; stopping and escalating is proper troubleshooting.

Clinical Use Tip

Unexpected patient weight change or fluid removal should be treated as a clinical concern first, not automatically as proof of machine failure.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Dialysis staff reported the Dialog+ repeatedly failed its fluid-balance check during preparation."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the drain hose was partially kinked behind the machine and restricting outflow."

Resolution

What action was taken.

Example:
"The drain hose was rerouted, the fluid-balance test completed successfully, and the machine passed functional verification before return to service."

Helpful Details to Include (If Known)

Final Thought

Protect the patient from unreliable fluid removal, verify settings and external hydraulic conditions before assuming internal failure, confirm performance with proper testing, escalate unresolved problems, and document the final result.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide