B. Braun Space Infusion Pump System

Air-in-Line Alarm or Air Detector False Alarm

On this page

Asset Type

Infusion Pump

Manufacturer

B. Braun

Model

Space Infusion Pump System

What This Guide Helps With

Troubleshooting Infusomat Space air-bubble alarms caused by actual air, improper tubing placement, incompatible sets, contamination, damaged tubing, or detector failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an air-in-line alarm while the affected pump remains responsible for an active patient infusion.

Expected outcome: The patient continues receiving prescribed therapy without exposure to air or interruption from an unreliable pump.

The manufacturer instructs users to disconnect the line from the patient before following prompts to prime an air bubble from the tubing.

2. Record the Exact Alarm Message

Determine whether the display reports:

Review the event history when available and record whether the alarm occurred during startup, priming, transport, bag replacement, or active delivery.

Expected outcome: The specific alarm type and operating conditions are documented before the pump is disturbed.

3. Inspect the Entire Administration Set for Actual Air

With the tubing disconnected from the patient, inspect from the fluid container through the pump and downstream tubing.

Look for:

Do not assume the alarm is false until the complete fluid path has been checked.

Expected outcome: No visible air, foam, or source of air entry remains in the administration set.

If air is present, have qualified clinical staff replace or reprime the set according to facility policy. Stop troubleshooting if this resolves the alarm.

4. Verify the Administration Set Is Approved and Undamaged

Confirm that the installed tubing is the correct B. Braun Infusomat Space-compatible administration set for the intended therapy.

Inspect the section passing through the pump for:

The Space system is designed to operate with dedicated Infusomat Space administration sets.

Expected outcome: A compatible, undamaged set is installed.

Replace a questionable set with a new approved set. If the alarm clears, remove the original disposable from use and document it as the probable cause.

5. Check Tubing Placement in the Air-Sensor Guide

Close the roller clamp and open the pump door only after the tubing is disconnected from the patient.

Verify that:

B. Braun specifically directs that the tubing be firmly pressed into the air-sensor guide and properly inserted into the sensors.

Expected outcome: The tubing lies naturally and completely within the detector channel.

Reload the set correctly and retest. If the alarm no longer occurs, stop troubleshooting.

6. Inspect the Air-Detector Area for External Contamination

Remove the administration set and visually inspect the air-sensor guide using adequate lighting.

Look for:

Clean only with manufacturer-approved products and methods. Do not scrape the sensor, introduce liquid into the pump, or use abrasive tools.

Expected outcome: The detector channel is clean, dry, unobstructed, and physically intact.

Allow the area to dry completely before testing. If contamination removal resolves the alarm, complete functional verification before returning the pump to service.

7. Test With a New Approved Administration Set

Install a new, fully primed, air-free compatible set using water or the facility-approved test fluid.

Expected outcome: The pump operates without an air alarm when a new, correctly loaded, air-free set is used.

If the replacement set works normally, the original tubing, loading condition, or fluid setup was likely responsible.

8. Compare the Pump With a Known-Good Pump

Using the same approved test setup, compare the affected pump with another verified Infusomat Space pump.

Determine whether:

Do not repeatedly move a disposable between pumps when contamination, medication exposure, or evidence preservation is a concern.

Expected outcome: The problem follows either the administration setup or the pump.

If multiple approved air-free sets work in another pump but repeatedly fail in the affected unit, suspect an air-detector, sensor-channel, calibration, connection, or internal electronic problem.

9. Review Configuration and Service History

Air-bubble size and accumulated-air alarm limits are configurable through the service program. Do not alter these limits solely to eliminate nuisance alarms.

Check for:

Compare configuration only with an approved reference pump or facility configuration record.

Expected outcome: The device has the authorized configuration, and recurring history or recent changes are identified.

Unauthorized or unexplained configuration differences require escalation rather than field adjustment.

10. Perform the Approved Air-Detector Functional Test

At the bench, perform the manufacturer-authorized air-detector test using the proper service procedure, test equipment, and controlled bubble volume.

Confirm that the pump:

Do not substitute an improvised bubble size or adjust detector calibration without the current B. Braun service documentation and required authorization.

Expected outcome: The air detector consistently recognizes the approved test condition without producing alarms during an air-free test.

If the pump fails either portion of the test, remove it from service.

If the Problem Persists

If multiple approved, properly primed administration sets have been tested, the sensor channel is clean, tubing is correctly loaded, configuration is verified, and the pump still produces false or inconsistent alarms, common external causes have been ruled out.

The problem is likely associated with the air detector, sensor alignment, internal connection, calibration, or control electronics.

The device should be:

Do not reduce the configured air-detection sensitivity to compensate for a suspected malfunction. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never clear an air alarm and restart an infusion without verifying the complete tubing path. Disconnect the line from the patient before priming, and transfer therapy before troubleshooting an unreliable detector.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported repeated “Air bubble alarm” messages on the B. Braun Infusomat Space despite no visible air in the administration tubing."

Cause

What was observed during troubleshooting.

Example:
"The administration set was not fully seated in the air-sensor guide, causing intermittent false air detection."

Resolution

What action was taken.

Example:
"Removed the pump from patient use, reloaded a new approved primed set correctly, cleaned the detector channel, and verified normal operation and air-alarm response during bench testing."

Helpful Details to Include (If Known)

Final Thought

Air-in-line complaints require a careful distinction between an actual air hazard, an administration-set problem, and a malfunctioning detector. Protect the patient first, inspect the entire external fluid path, verify correct set loading, and escalate any pump that cannot reliably pass controlled testing. Complete CCR documentation supports patient safety and future failure analysis.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide