B. Braun Space Infusion Pump System

Battery Not Charging or Rapid Battery Depletion

On this page

Asset Type

Infusion Pump

Manufacturer

B. Braun

Model

Space Infusion Pump System

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting B. Braun Space Infusion Pumps that fail to charge properly or display rapid battery depletion. Common causes are often external, such as docking issues, power supply problems, or battery seating errors, rather than internal battery failure.

Step-by-Step Troubleshooting

Verify Power Source

Inspect Docking Station Connection

Check Power Cable and Adapters

Confirm Battery Seating

Review Pump Indicators

Perform Battery Calibration / Pump Reset

Swap Battery (if available)

If the Problem Persists

External factors have been ruled out. The battery or internal charging circuitry may be failing.

Knowing when to stop is proper troubleshooting and ensures patient safety.

Clinical Use Tip

Never attempt patient infusion with a pump showing charging errors.

Move the patient to another verified pump before troubleshooting.

Only perform troubleshooting when the device is not actively infusing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump battery will not charge and drains rapidly; pump disconnected from dock.”

Cause

What was observed during troubleshooting.

Example:
Verified power source, docking station, and cables; battery correctly seated. Pump battery continues to deplete quickly, suggesting possible internal battery or charging circuit failure.

Resolution

What action was taken.

Example:
Pump removed from service, labeled Out of Service, and sent to biomedical repair for evaluation.

Helpful Details to Include

Final Thought

Patient safety and logical troubleshooting go hand-in-hand. Start with external and easily verified checks, escalate appropriately, and document all findings. This prevents unnecessary downtime, ensures accurate diagnosis, and maintains compliance.

That is successful troubleshooting.