B. Braun Space Infusion Pump System

Incorrect Flow Rate / Rate Deviation Alarms

On this page

Asset Type

Infusion Pump

Manufacturer

B. Braun

Model

Space Infusion Pump System

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting B. Braun Space Infusion Pumps that trigger incorrect flow rate or rate deviation alarms. These alarms indicate the pump is delivering fluid faster or slower than programmed. In many cases, the cause is external or setup-related rather than internal pump failure, such as tubing misplacement, incorrect cassette, or occlusion in the line.

Step-by-Step Troubleshooting

Verify Pump Programming

Why: User entry errors are a common cause of flow rate alarms.

Check IV Line and Tubing

Why: Misplaced tubing or blocked lines can alter flow rates.

Ensure Proper Cassette Seating

Why: Improper cassette placement can trigger rate deviation alarms.

Check for Air or Bubbles

Why: Air in the line can cause inconsistent flow and alarms.

Verify Pump Orientation

Why: Some pumps have flow sensors sensitive to tilt or movement.

Inspect Power Source and Battery

Why: Voltage fluctuations can affect pump performance and alarms.

Reset the Pump if Safe

Why: Software glitches can sometimes cause false rate deviation alarms.

If the Problem Persists

External causes have been ruled out. The device should be:

Knowing when to escalate avoids unnecessary risk to patients and prevents further device damage.

Clinical Use Tip

Do not troubleshoot on an active patient. Move the patient to a different pump if an alarm persists. Ensure infusion continuity and patient safety before any pump testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump displays rate deviation alarm; infusion rate does not match programmed value.”

Cause

What was observed during troubleshooting.

Example:
“IV tubing kinked and cassette not fully seated; external check completed.”

Resolution

What action was taken.

Example:
“Corrected tubing placement, reseated cassette, and test infusion ran within acceptable limits. Pump returned to service.”

Helpful Details to Include

Final Thought

Patient safety is the priority. Start with logical, external checks before assuming internal failure. Accurate documentation ensures clarity for future troubleshooting and highlights proper escalation procedures.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide