B. Braun Space Infusion Pump System

Keypad Not Responding

On this page

Asset Type

Infusion Pump

Manufacturer

B. Braun

Model

Space Infusion Pump System

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting B. Braun Space Infusion Pumps when the keypad buttons fail to respond to user input. Symptoms may include buttons not registering or intermittent response when adjusting settings.

Many of these issues are caused by simple external factors such as contamination, moisture, temporary software lockups, or power-related problems rather than internal hardware failure.

Step-by-Step Troubleshooting

Confirm the Pump Is Powered and Fully Booted

Check for Active Alarms or Lock Screens

Inspect the Keypad for Contamination

Test Individual Buttons

Clean the Keypad Surface

Power Cycle the Pump

Test Operation Off the Dock

Check for Physical Damage

Test Menu Navigation

If the Problem Persists

If the keypad remains unresponsive after these external checks, common external causes have been ruled out. The issue is likely due to an internal failure such as keypad membrane failure, rotary encoder failure, or a user interface board fault.

At this point the device should be:

Recognizing when internal repair is required is an important part of proper troubleshooting.

Clinical Use Tip

Never troubleshoot an infusion pump that is actively connected to a patient.

If keypad or control issues occur during therapy, transfer the infusion to another pump, ensure the patient’s therapy continues safely, and remove the malfunctioning pump for evaluation. Patient therapy should never depend on a device with unreliable controls.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
Nursing staff reported the infusion pump keypad was not responding when attempting to adjust infusion settings.

Cause

What was observed during troubleshooting.

Example:
Testing confirmed keypad inputs were intermittent. External contamination ruled out; likely internal keypad/encoder failure.

Resolution

What action was taken.

Example:
Pump removed from service and labeled Out of Service. Device sent for repair and replacement pump issued to the unit.

Helpful Details to Include

Final Thought

Interface failures on infusion pumps can significantly impact safe medication delivery. Clinical Engineering should always begin with external causes such as contamination, software lockups, or power issues before assuming hardware failure. When input controls remain unreliable after these checks, removing the device from service protects both patients and clinical staff. Clear documentation ensures the issue is properly tracked and resolved.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide