B. Braun Space Infusion Pump System

Pump Will Not Start or Resume Infusion

On this page

Asset Type

Infusion Pump

Manufacturer

B. Braun

Model

Space Infusion Pump System

What This Guide Helps With

Troubleshooting start or resume failures caused by active alarms, incomplete programming, tubing or syringe loading, closed clamps, power issues, or accessory problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a pump that will not start or resume while the patient depends on it for medication delivery.

Expected outcome: Prescribed therapy continues safely without depending on the affected pump.

2. Identify the Exact Start Failure

Determine whether:

Record the complete displayed message and any alarm tone before clearing it.

Expected outcome: The failure is clearly identified as a programming, loading, alarm, control, power, or device problem.

3. Check for an Active Alarm or Prompt

Review the display for unresolved conditions such as:

A B. Braun Space pump may prevent infusion until the displayed condition has been acknowledged and corrected.

Correct only conditions that can be safely verified externally. Do not repeatedly silence or clear an alarm without identifying its cause.

Expected outcome: The alarm clears and the pump returns to a ready-to-start state.

If the infusion starts normally, verify operation and stop troubleshooting.

4. Verify the Infusion Is Fully Programmed

Confirm that all required values have been entered and accepted, including the applicable:

Check for a confirmation screen, flashing value, unanswered question, or soft-limit notification that still requires acknowledgment.

Do not alter prescribed settings solely to make the pump start. Clinical staff must verify all therapy parameters.

Expected outcome: The programmed infusion is complete, accepted, and ready to begin.

5. Confirm the Pump Is Not in Standby or Pause Mode

Check the display for standby, stop, pause, or delayed-start status.

Expected outcome: The status changes from stopped or paused to an active infusion indication.

If the pump begins infusing and remains stable, the problem is resolved.

6. Inspect the Power Source

Confirm that:

Test the wall outlet with an approved electrical tester or known-good device. When available, test the pump with a known-good compatible power supply or docking position.

The Space system may be operated individually or integrated into a SpaceStation configuration, making both pump and station connections relevant.

Expected outcome: The pump receives stable power without battery or docking-related interruptions.

7. Inspect the Administration Set on an Infusomat Space Pump

For a large-volume Infusomat Space pump:

The Infusomat Space system uses dedicated administration sets and accessories.

When safe, reload the set according to facility training. Use a new compatible set if the installed set is damaged or questionable.

Expected outcome: The pump recognizes the set, the door closes normally, and no loading or occlusion alarm prevents startup.

8. Inspect the Syringe on a Perfusor Space Pump

For a Perfusor Space syringe pump:

Do not force the syringe holder or drive mechanism.

Expected outcome: The syringe is recognized, mechanically secured, and accepted by the pump.

9. Check for an External Occlusion

Trace the entire infusion pathway from the fluid source or syringe to the patient connection.

Look for:

Clinical staff should evaluate the vascular access device. Clinical Engineering should not flush or manipulate the patient’s access.

Expected outcome: The external flow path is unobstructed and the pump starts without an immediate pressure alarm.

10. Check Air-Detection Conditions

For an Infusomat Space pump, inspect the administration set near the air detector.

Never bypass an air alarm or alter the air-detection configuration to force the infusion to start.

Expected outcome: The air condition is corrected and the pump permits startup.

11. Inspect Controls and Display Response

Press the start control once and observe whether:

Clean visible contamination from the control surface using an approved method. Do not allow liquid to enter seams or openings.

If the start key is physically damaged, intermittent, or unresponsive while other controls work, remove the pump from service.

Expected outcome: The controls respond consistently and the start command is recognized.

12. Test the Pump Away From the Docking System

When the pump is installed in a SpaceStation:

Do not scrape, bend, or probe electrical contacts.

Expected outcome: The test identifies whether the failure follows the pump or remains with the station position.

If the problem remains with one station position, remove that docking position or station from service for evaluation.

13. Perform a Controlled Restart

Only after the patient has been transferred and infusion information has been preserved:

Do not repeatedly reboot a pump that displays a technical error or is associated with a suspected medication event.

Expected outcome: The pump completes startup, accepts programming, and responds normally.

14. Conduct a Non-Patient Functional Test

Use facility-approved test supplies and procedures.

Do not return the pump to service based only on the appearance of normal screen operation.

Expected outcome: The pump reliably starts, stops, resumes, and delivers within the facility’s acceptance criteria.

If the pump passes all testing, document the findings and return it to service according to departmental policy.

If the Problem Persists

If the pump still will not start or resume after programming, loading, power, alarm, tubing, syringe, control, and docking checks, common external causes have been ruled out.

The problem may involve the start control, door or syringe-detection system, drive mechanism, pressure or air sensor, power circuit, software, or another internal component.

The device should be:

Do not bypass sensors, defeat alarms, force mechanical components, or perform unauthorized internal repairs.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a start or resume failure while a patient depends on the affected pump. Move the therapy to another verified device before manipulating the tubing, syringe, door, power connection, or programming.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the B. Braun Space infusion pump accepted programming but would not start or resume the infusion."

Cause

What was observed during troubleshooting.

Example:
"Inspection found that the administration set was incorrectly seated in the Infusomat Space tubing channel, preventing the door from fully recognizing the loaded set."

Resolution

What action was taken.

Example:
"Removed the pump from patient use, correctly installed a new compatible administration set, completed start-stop-resume and delivery testing, and returned the pump to service after all tests passed."

Helpful Details to Include (If Known)

Final Thought

A pump that will not start or resume can interrupt critical medication delivery. Protect the patient first, verify external causes logically, test the complete infusion setup, and escalate when reliable operation cannot be demonstrated. Clear CCR documentation supports both patient safety and effective repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide