Baxter Novum IQ

Air-in-Line Detector Errors

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

Novum IQ

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting air-in-line detector errors on the Novum IQ infusion pump. These alarms typically indicate air bubbles detected in the fluid path, but are often caused by setup issues, tubing problems, or sensor contamination before assuming internal detector failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Alarm Message and Behavior

Inspect the IV Tubing for Air Bubbles

Check for Improper Priming

Verify Correct Tubing Type and Setup

Inspect the Air-in-Line Sensor Area

Check for Kinks or Irregular Flow

Swap Tubing Set

Test with a Known Good Setup

Power Cycle the Pump

If the Problem Persists

If air-in-line alarms continue after verifying proper priming, tubing, and sensor cleanliness, the issue is likely internal (e.g., faulty air-in-line sensor or detection circuitry).

Knowing when to stop prevents unnecessary risk and ensures patient safety.

Clinical Use Tip

Do not troubleshoot air-in-line alarms while the pump is actively infusing on a patient. Always transfer the patient to a properly functioning pump before continuing troubleshooting. Air-in-line conditions can pose a serious risk to the patient if not addressed immediately.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nursing reported repeated air-in-line alarms during infusion despite no visible air in tubing.”

Cause

What was observed during troubleshooting.

Example:
“Observed no air in line; sensor area had visible residue causing intermittent false detection.”

Resolution

What action was taken.

Example:
“Cleaned sensor area, replaced tubing, and tested with known good setup. Alarm cleared and unit returned to service.”

Helpful Details to Include

Final Thought

Air-in-line alarms should always be taken seriously, but most causes are external and correctable with careful inspection and proper setup. A structured troubleshooting approach helps quickly rule out user and setup errors while ensuring patient safety. When the issue points to internal failure, timely escalation and clear documentation are essential.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide