Baxter Novum IQ

Battery or Power Module Failure

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

Novum IQ

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where the Novum IQ infusion pump does not power on, intermittently loses power, fails to charge, or displays battery-related alarms. The focus is on verifying external power sources, connections, and battery condition before assuming internal power module failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify AC Power Source

Inspect the Power Cord and Adapter

Check Power Indicator Lights

Attempt Power-On with AC Connected

Check Battery Status on Screen (if accessible)

Inspect Battery Compartment (External Only)

Swap with Known Good Power Cord or Battery (if available)

Check for Intermittent Power Loss

Listen and Observe During Charging

If the Problem Persists

If all external power sources, cords, and battery checks have been verified and the issue remains, the problem is likely internal (battery failure, charging circuit issue, or power module failure).

Recognizing when the issue is beyond external troubleshooting is critical to safe and effective Clinical Engineering practice.

Clinical Use Tip

Do not troubleshoot power issues on a device connected to a patient. Transition the patient to a properly functioning infusion pump before removing this device from service.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump would not power on and displayed battery failure alarm during use.”

Cause

What was observed during troubleshooting.

Example:
“Battery not holding charge and device only powered on when connected to AC. Verified with known good outlet and power cord.”

Resolution

What action was taken.

Example:
“Removed pump from service and sent for battery replacement and internal power module evaluation.”

Helpful Details to Include

Final Thought

Power-related issues should always be approached from the outside in—starting with the simplest, most accessible causes. Ensuring reliable power is critical for patient safety, especially with infusion therapy. Knowing when to escalate protects both the patient and the equipment, while clear documentation supports ongoing reliability and accountability.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide