Baxter Novum IQ

Occlusion Alarm Troubleshooting

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

Novum IQ

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting occlusion alarms on the Novum IQ infusion pump. These alarms typically indicate resistance in the fluid path. The goal is to rule out common external causes such as tubing issues, clamps, or setup problems before considering internal sensor or pump mechanism failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Alarm Condition

Check for Closed or Partially Closed Clamps

Inspect the IV Tubing for Kinks or Obstructions

Verify Proper Tubing Installation in the Pump

Check the IV Site (If Applicable)

Inspect the Fluid Bag and Upstream Flow

Verify Correct Administration Set Type

Power Cycle the Pump

Test with a Known Good Setup

If the Problem Persists

If all external causes have been ruled out and the occlusion alarm continues:

Knowing when to stop external troubleshooting is critical to safe and effective Clinical Engineering practice.

Clinical Use Tip

Do not troubleshoot occlusion alarms on an active patient infusion. Always have clinical staff transition the patient to a functioning pump before continuing troubleshooting. This ensures uninterrupted therapy and patient safety.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nursing reported frequent occlusion alarms during infusion despite proper setup.”

Cause

What was observed during troubleshooting.

Example:
“Kinked tubing found under bed rail causing intermittent downstream occlusion.”

Resolution

What action was taken.

Example:
“Repositioned tubing, verified proper flow, and confirmed pump operation without further alarms.”

Helpful Details to Include

Final Thought

Occlusion alarms are often caused by simple, external restrictions in the fluid path. A structured, step-by-step approach ensures quick resolution while protecting patient safety. When external causes are eliminated, timely escalation prevents unnecessary downtime and ensures reliable device performance. Clear documentation supports both patient care and equipment history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide