Baxter PrismaFlex

Alarm: Access Extremely Negative

On this page

Asset Type

Continuous Renal Replacement Therapy (CRRT)

Manufacturer

Baxter

Model

PrismaFlex

What This Guide Helps With

This guide helps troubleshoot the Baxter PrismaFlex Access Extremely Negative alarm. This alarm can occur when access pressure becomes excessively negative due to catheter restriction, access-line obstruction, excessive blood flow demand, patient or catheter position changes, disposable set issues, or pressure-sensing failure.

The condition may interrupt CRRT therapy, stop or slow the blood pump, or indicate poor blood withdrawal from the patient access catheter.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient care is maintained safely while the access-pressure issue is evaluated.

Why it matters: An unresolved access pressure alarm can indicate poor blood withdrawal from the catheter and may interrupt therapy or contribute to circuit complications.

Verify the Reported Alarm and Behavior

Expected: The alarm is confirmed and linked to access pressure becoming excessively negative.

Why it matters: This helps separate a true access-pressure problem from a one-time nuisance alarm or operator misinterpretation.

Allow the Self-Clearing Attempt When Appropriate

Expected: A transient access-pressure drop clears and does not immediately return.

Why it matters: Brief patient movement or catheter position changes can temporarily create excessive negative access pressure.

Check the Access Line for External Obstruction

Expected: The access line is open, unkinked, properly routed, and free of tension.

Why it matters: A clamped or restricted access line can cause the pump to pull harder against the access source, making access pressure extremely negative.

Check for Patient Movement or Catheter Position Effects

Expected: Access pressure improves when the patient and catheter are positioned appropriately.

Why it matters: A malpositioned catheter can intermittently occlude against the vessel wall and create very negative access pressure.

Have Clinical Staff Evaluate the Access Catheter

Expected: The catheter provides adequate blood withdrawal for the ordered blood flow rate.

Why it matters: A clotted, malpositioned, or partially occluded catheter is a common cause of extremely negative access pressure.

Compare Blood Flow Rate to Access Capability

Expected: Access pressure improves if the blood flow demand was exceeding what the catheter could provide.

Why it matters: If the blood pump demand exceeds the catheter’s ability to supply blood, the system may generate excessive negative access pressure.

Inspect the Pressure Pod and Set Interface

Expected: The disposable set and access pressure pod are correctly installed with no visible leakage or damage.

Why it matters: The PrismaFlex relies on accurate pressure sensing to detect access pressure changes. A poorly seated or compromised pod can cause false or unstable pressure readings.

Determine Whether the Issue Is Access-Related or Device-Related

Expected: Troubleshooting narrows the issue to an accessory, disposable set, catheter, blood flow setting, or device pressure-sensing problem.

Why it matters: The device should not be sent for repair until external and access-related causes have been reasonably ruled out.

Escalate for Pressure-Sensing Evaluation if Needed

Expected: Faulty pressure-sensing hardware is identified, calibrated if appropriate, or replaced if deviation remains.

Why it matters: A failed or drifting access pressure sensor can cause repeated false or unresolved Access Extremely Negative alarms.

If the Problem Persists

If the alarm continues after the access line, patient position, catheter function, blood flow rate, pressure pod, and disposable set have been checked, external and common access-related causes have been ruled out.

A persistent Access Extremely Negative alarm may indicate a pressure-sensing fault, ARPS leakage, calibration issue, or internal device failure.

Remove the PrismaFlex from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot an unresolved access pressure alarm on an active patient longer than clinically appropriate. Move the patient to a backup device or alternate therapy plan first, and involve nursing or the provider for catheter assessment, flushing, repositioning, or blood-flow adjustments.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PrismaFlex removed from CRRT use due to Access Extremely Negative alarm. Staff reported alarm during treatment with access pressure becoming excessively negative and blood pump interruption."

Cause

What was observed during troubleshooting.

Example:
"Access line and external routing were inspected. No clamp or kink found. Alarm persisted after patient repositioning and blood flow reduction. Catheter issue ruled out by clinical staff. Suspect access pressure-sensing or ARPS circuit fault."

Resolution

What action was taken.

Example:
"Device removed from service and labeled Out of Service. Sent for bench evaluation of access pressure sensor, ARPS circuit leakage, and pressure-sensor calibration. Repair required before return to patient use."

Helpful Details to Include (If Known)

Final Thought

The Access Extremely Negative alarm should be treated as a real blood access concern first, not immediately as a device failure. Start with patient safety, line patency, catheter position, and blood flow demand. Once those are ruled out, persistent alarms justify pressure-sensing evaluation and service escalation. Good documentation connects the displayed alarm to the clinical symptom, the checks performed, and the final disposition.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide