Baxter PrismaFlex

Alarm: Air in Blood

On this page

Asset Type

Continuous Renal Replacement Therapy (CRRT)

Manufacturer

Baxter

Model

PrismaFlex

What This Guide Helps With

This guide helps troubleshoot the Baxter PrismaFlex Air in Blood alarm when air, foam, leakage, incomplete priming, or return-line installation issues affect the blood return path. This alarm can interrupt CRRT therapy and must be treated as a patient safety concern because air in the return path may create a serious risk to the patient.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Stop and Inspect Before Continuing

4. Check the Blood Access and Disposable Set for Leaks

5. Verify Return Line Placement in the Air Detector

6. Inspect for Visible Air or Foam

7. Follow the On-Screen Air Removal Process

8. Determine Whether the Set Was Fully Primed

9. Replace the Set if Air Is Present Throughout the Circuit

10. Confirm Alarm Resolution

If the Problem Persists

If the Air in Blood alarm continues after leaks and disconnections have been ruled out, return line installation has been verified, visible air or foam has been removed, the set has been properly primed, and a new disposable set has been loaded if needed, then external and common disposable-set causes have been ruled out.

Persistent Air in Blood alarms may indicate a device-related issue such as air detector failure, sensor malfunction, or related detection circuitry failure.

Remove the PrismaFlex from service, label it Out of Service, and send it for Clinical Engineering bench evaluation or repair.

Clinical Use Tip

Never troubleshoot an unresolved Air in Blood alarm on an active patient beyond the device-guided and facility-approved process. Move the patient to a backup device or alternate therapy support first if treatment must continue, the set must be changed, or the alarm cannot be cleared safely.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PrismaFlex CRRT reported Air in Blood alarm during treatment. Staff observed air or foam in the return path, and therapy could not continue until the alarm was corrected."

Cause

What was observed during troubleshooting.

Example:
"Air was detected in the blood return path. External inspection found possible disposable-set air source, return-line installation issue, incomplete priming, leakage, or air/foam in tubing. If unresolved after set replacement and proper setup, likely air detector or sensor-related failure."

Resolution

What action was taken.

Example:
"Inspected blood access and disposable set for leakage or disconnection. Verified return line placement in the air detector. Checked return tubing and deaeration chamber for visible air or foam. Followed on-screen air removal process using negative return pressure and RELEASE CLAMP when prompted. If air remained throughout the set, treatment was disconnected and a new set was loaded and primed. Device returned to service only after alarm cleared and no visible air remained. If alarm persisted, device was removed from service for bench evaluation."

Helpful Details to Include (If Known)

Final Thought

The Air in Blood alarm should always be treated as a patient safety issue first. Logical troubleshooting begins with the blood path, disposable set, tubing placement, visible air, and priming status before suspecting device failure. If the alarm cannot be cleared safely, removing the device from service and documenting the findings protects the patient and supports proper repair follow-up.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide