On this page
Asset Type
Continuous Renal Replacement Therapy (CRRT)
Manufacturer
Model
What This Guide Helps With
This guide helps troubleshoot a Baxter PrismaFlex Blood Leak Detected alarm caused by possible blood in the effluent path, tubing-position issues, air bubbles, detector contamination, discolored plasma, filter membrane leakage, or blood leak detector failure.
A true blood leak can indicate filter failure and possible patient blood loss into the effluent line, so this alarm must be handled as a patient-safety condition until proven otherwise.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Respond to the alarm immediately and follow facility CRRT escalation procedures.
- Do not ignore, silence repeatedly, or bypass a Blood Leak Detected alarm.
- If the patient is actively connected, coordinate with nursing and clinical staff before handling the set, effluent line, detector area, or blood return process.
- If therapy cannot safely continue, move the patient to an alternate CRRT device or therapy plan according to facility policy.
Verify the Reported Alarm and Behavior
- Confirm the exact alarm displayed is Blood Leak Detected.
- Document whether the alarm occurred during active treatment, immediately after setup or priming, after patient connection, during TPE or plasma-related therapy, or intermittently with effluent-line movement.
- Check whether the alarm returns after normalization or line reinsertion.
- Observe whether the effluent line appears clear, pink, red, cloudy, discolored, or visually normal.
Inspect the Effluent Line at the Blood Leak Detector
- Visually inspect the effluent line where it passes through the blood leak detector.
- Look for an air bubble in the detector window.
- Verify the tubing is fully seated and correctly routed through the detector.
- Check for tubing that is twisted, stretched, pinched, pulled out of position, or routed incorrectly.
- Correct any obvious tubing-position issue and reinstall the effluent line properly in the detector.
Check for Air Bubbles at the Detector
- If an air bubble is visible at or near the blood leak detector, work with clinical staff to clear the bubble according to facility procedure and PrismaFlex operating practice.
- Do not shake, pull, or aggressively manipulate the line while the patient is connected.
- Confirm the detector path is consistently filled with effluent and that no bubble remains in the optical detection area.
Check for Moisture, Liquid, or Debris in the Detector Area
- Inspect the blood leak detector area for effluent drips, cleaning-solution residue, dried fluid, dust, debris, or moisture.
- Check whether the outside of the tubing is wet.
- Look for cracked or leaking tubing near the detector.
- Clean and dry the external detector area according to approved facility and manufacturer cleaning practices.
- Do not force liquid into the detector.
Reinsert the Line and Normalize the Blood Leak Detector if Appropriate
- After correcting tubing placement, bubbles, or external contamination, reinsert the effluent line correctly.
- During treatment, normalize the blood leak detector after reinsertion only when clinically appropriate and consistent with PrismaFlex procedure and facility policy.
- Verify whether the alarm clears and whether treatment continues without repeated Blood Leak Detected alarms.
Evaluate for True Filter Membrane Leakage
- If the alarm persists or the effluent appears visibly blood-tinged, treat the condition as a possible true blood leak.
- Notify clinical staff and follow facility protocol for stopping therapy, changing the set, and managing the patient’s blood return status.
- If filter membrane leakage is suspected, the set should be changed.
- Send an effluent sample for cell count when required by clinical policy.
Consider Therapy-Specific False Triggers
- If the unit is being used for TPE or a therapy where plasma discoloration may be present, verify whether effluent or plasma color could be affecting the detector reading.
- Do not assume the alarm is false without clinical review.
- Continue treating the alarm as a patient-safety issue until tubing, detector condition, therapy factors, and filter integrity have been evaluated.
Test With a New Set if Clinically Appropriate
- If the set is changed due to suspected filter membrane leakage or repeated alarms, monitor whether the Blood Leak Detected alarm recurs with the new set.
- If the alarm clears with a new set, the issue was likely set, filter, fluid-path, or tubing related.
- If the alarm recurs with a new set and a clean detector path, suspect a device-side blood leak detector issue.
Determine Whether the Problem Is Accessory, Set, or Device Related
- If the alarm cleared after removing a bubble, the likely cause was air at the detector.
- If the alarm cleared after reinstalling tubing, the likely cause was incorrect effluent-line position.
- If the alarm cleared after cleaning and drying the detector path, the likely cause was detector-path contamination.
- If the alarm cleared after set replacement, the likely cause was set, filter, or fluid-path related.
- If the alarm recurs with a new set and clean detector path, the device blood leak detector may be faulty.
Stop When Resolved
- If the alarm clears after correcting the effluent line, removing bubbles, cleaning the detector path, normalizing appropriately, or replacing the set, verify stable operation before returning the PrismaFlex to clinical use.
- Document the action taken, final alarm status, whether a set was changed, and whether the device remained in service or was removed for repair.
If the Problem Persists
If the Blood Leak Detected alarm continues after the effluent line is correctly installed, bubbles are cleared, the detector path is clean and dry, and a new set has been tried, external and common causes have been ruled out.
Persistent Blood Leak Detected alarms strongly indicate a device-side blood leak detector issue or related sensing fault.
Remove the PrismaFlex from service, label it Out of Service, and send it for repair or bench evaluation.
Clinical Engineering should check the blood leak detector in Service Diagnose. If the blood leak detector fails testing or cannot reliably detect normal conditions, replace the blood leak detector as indicated.
Clinical Use Tip
Never troubleshoot a Blood Leak Detected alarm casually on an active patient. Treat it as a possible true blood leak until tubing, detector condition, therapy factors, and filter integrity are evaluated.
Move the patient to a backup device or alternate therapy plan first when continued CRRT support is required and the alarm cannot be confidently resolved.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"PrismaFlex reported Blood Leak Detected alarm during treatment. Staff reported alarm at the effluent line/blood leak detector area with concern for possible blood in the effluent path."
Cause
What was observed during troubleshooting.
Example:
"Effluent line inspected for blood leak detector false-trigger causes. Tubing position, air bubble at detector, detector-path contamination, moisture, and set/filter condition evaluated. Alarm suspected to be caused by incorrect effluent-line seating and air bubble at detector."
Resolution
What action was taken.
Example:
"Removed device from patient dependency during evaluation. Reinstalled effluent line correctly, cleared visible bubble from detector area with clinical staff, verified detector path was clean and dry, normalized detector after reinsertion per procedure, and confirmed alarm cleared without recurrence. Device returned to service after successful functional verification."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
The Blood Leak Detected alarm must be approached with patient safety first. A true blood leak can indicate filter membrane failure, while false triggers may come from bubbles, tubing position, contamination, plasma discoloration, or detector malfunction.
Clinical Engineering should rule out simple external causes first, support clinical escalation when filter leakage is suspected, and remove the device from service when the alarm persists after set and detector-path checks. Clear CCR documentation protects the patient, the clinical team, and the repair process.
That is successful troubleshooting.