Baxter PrismaFlex

Alarm: Filter Clotted

On this page

Asset Type

Continuous Renal Replacement Therapy (CRRT)

Manufacturer

Baxter

Model

PrismaFlex

What This Guide Helps With

This guide helps troubleshoot the Baxter PrismaFlex Alarm: Filter Clotted, which may indicate excessive filter pressure drop, suspected clotting, restricted blood flow, anticoagulation delivery problems, excessive ultrafiltration demand, or pressure measurement issues during CRRT therapy.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Translate the Alarm Into the Real-World Condition

Check Blood-Flow Lines for Clamps or Kinks

Assess for Visible or Suspected Circuit Clotting

Confirm Anticoagulation Status With Clinical Staff

Check Syringe Pump or PBP Pump Delivery

Evaluate Ultrafiltration and Fluid Removal Demands

Check for Pressure Measurement Issues

Determine Whether the Issue Is Circuit-Related or Device-Related

Stop When Resolved

If the Problem Persists

If the Filter Clotted alarm continues after external causes, circuit issues, therapy settings, and anticoagulation delivery have been checked, remove the PrismaFlex from service.

At this point, external tubing restriction, disposable set issues, therapy setting problems, visible circuit clotting, anticoagulation interruption, and pressure pod setup concerns should be considered ruled out as practical bedside causes.

Persistent or repeated Filter Clotted alarms after a new set may indicate pressure measurement failure, syringe pump malfunction, PBP pump malfunction, or another device-side fault.

Clinical Use Tip

Never troubleshoot a suspected clotted CRRT circuit while relying on it for active patient therapy. Coordinate with the bedside team, move the patient to a backup device or alternate therapy plan first, and change the set when clotting is confirmed or strongly suspected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PrismaFlex removed from service due to repeated Filter Clotted alarm during CRRT treatment. Staff reported excessive filter pressure drop and clotting advisory or TMP caution conditions."

Cause

What was observed during troubleshooting.

Example:
"Inspection found no external tubing clamp or kink. Alarm recurred after user checks, indicating likely circuit clotting or possible pressure measurement, syringe pump, or PBP pump delivery issue."

Resolution

What action was taken.

Example:
"Device removed from patient use and labeled Out of Service. Recommended bench evaluation of pressure sensing function and syringe/PBP pump operation before return to service."

Helpful Details to Include (If Known)

Final Thought

The Filter Clotted alarm should be approached as both a patient safety issue and a circuit performance issue. Start with simple external checks, confirm whether clotting or restricted flow is present, review anticoagulation and ultrafiltration demands, and escalate when repeated alarms suggest pressure sensing or pump failure. Clear documentation helps separate disposable set issues from true device problems.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide