Baxter PrismaFlex

Alarm: Filter Extremely Positive

On this page

Asset Type

Continuous Renal Replacement Therapy (CRRT)

Manufacturer

Baxter

Model

PrismaFlex

What This Guide Helps With

This guide helps troubleshoot the Baxter PrismaFlex Filter Extremely Positive alarm. This alarm may occur when filter pressure reaches or exceeds 450 mmHg and can be associated with restricted blood flow, filter clotting, high return pressure, catheter issues, tubing restriction, or pressure-sensing failure.

The alarm should be treated as a potentially serious CRRT pressure condition because excessive filter pressure can affect therapy delivery, circuit safety, and patient care.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Expected outcome: Patient care is protected before device or circuit troubleshooting begins.

2. Verify the Reported Alarm and Behavior

Expected outcome: The alarm is confirmed and the pressure condition is clearly documented.

3. Translate the Alarm Into the Real-World Condition

Expected outcome: Troubleshooting stays focused on pressure, restriction, clotting, catheter performance, and sensing accuracy.

4. Check for Clamped or Kinked Lines

Expected outcome: Lines are open, routed correctly, and free of visible restriction.

5. Confirm Circuit Routing and Set Position

Expected outcome: The set is installed correctly with no external compression or misrouting.

6. Check for High Return Pressure

Expected outcome: Return pressure is understood as either normal or contributing to the filter pressure alarm.

7. Lower Blood Flow Rate if Clinically Appropriate

Expected outcome: Reduced blood flow may reduce pressure enough to continue safe therapy if the circuit and access are otherwise acceptable.

8. Inspect for Developing Filter Clotting

Expected outcome: Filter condition is assessed before assuming a machine failure.

9. Relieve Excess Pressure if the System Allows

Expected outcome: Excess pressure is relieved and the alarm clears without recurring.

10. Stop When Resolved

Expected outcome: The PrismaFlex remains in service with stable pressures and no recurring Filter Extremely Positive alarm.

11. Change the Set if Operator Correction Does Not Clear the Alarm

Expected outcome: Therapy is restarted only with a safe, functional circuit.

12. Evaluate for Repeated Alarms After Circuit Correction

Expected outcome: Repeated alarms are narrowed toward device pressure sensing rather than disposable set or access problems.

13. Perform Bench Evaluation if Device Failure Is Suspected

Expected outcome: Pressure sensing is verified before the device is returned to service.

If the Problem Persists

If external restrictions, clamps, tubing kinks, catheter issues, blood flow settings, excessive pressure, and circuit clotting have been ruled out, persistent Filter Extremely Positive alarms strongly suggest a pressure-sensing or internal device problem.

Remove the PrismaFlex from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first when therapy cannot safely continue on the alarming PrismaFlex.

Coordinate with nursing and the provider before changing flow conditions, stopping therapy, releasing pressure, returning blood, or changing the set. Avoid relying on a PrismaFlex with repeated unresolved pressure alarms.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PrismaFlex reported Filter Extremely Positive alarm during CRRT. Filter pressure reached or exceeded 450 mmHg. Staff reported repeated pressure alarms and concern for restricted blood flow or filter clotting."

Cause

What was observed during troubleshooting.

Example:
"Inspection found kinked tubing between the filter pressure pod and filter, causing elevated filter pressure. Alarm cleared after tubing was rerouted and pressure readings stabilized."

Resolution

What action was taken.

Example:
"Corrected tubing route, verified no clamps or kinks remained, confirmed filter pressure returned to acceptable range, and monitored for recurring alarms. Device remained in service after clinical staff confirmed therapy could continue."

Helpful Details to Include (If Known)

Final Thought

The Filter Extremely Positive alarm should be treated as a real pressure condition until proven otherwise. Start with patient safety, then check for clamps, kinks, circuit restriction, catheter issues, high return pressure, and filter clotting before suspecting the machine. If the alarm persists after circuit and clinical causes are addressed, escalation to pressure-sensor or ARPS evaluation is appropriate. Clear CCR documentation protects both patient care and the repair process.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide