Baxter PrismaFlex

Alarm: HP Cartridge Clotted

On this page

Asset Type

Continuous Renal Replacement Therapy (CRRT)

Manufacturer

Baxter

Model

PrismaFlex

What This Guide Helps With

This guide helps troubleshoot the Baxter PrismaFlex HP Cartridge Clotted alarm. This alarm may be associated with cartridge pressure drop, hemopurification cartridge clotting, blood-flow restriction, anticoagulation-delivery problems, or pressure measurement issues.

A clotted or restricted HP cartridge can affect CRRT therapy delivery, increase circuit pressure, reduce treatment effectiveness, and may create patient safety concerns if the circuit condition prevents safe continuation of therapy.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Inspect the Blood-Flow Lines for Simple Restrictions

Check for Visible Cartridge or Circuit Clotting

Review Anticoagulation Status With Clinical Staff

Confirm the Anticoagulant Delivery Path

Evaluate Blood Flow Setting With the Care Team

Do Not Continue With a Suspected Clotted Cartridge

Perform Device Self-Test if Pressure Measurement Failure Is Suspected

Check Syringe Pump Function if Anticoagulant Delivery Failed

Isolate the Likely Cause

Stop When Resolved

If the Problem Persists

If the HP Cartridge Clotted alarm continues after external restrictions, cartridge clotting, anticoagulation setup, and syringe delivery issues have been addressed, remove the PrismaFlex from service.

Label the device Out of Service and send it for repair or bench evaluation.

At that point, external causes have been ruled out, and persistent alarms may indicate pressure measurement failure, syringe pump malfunction affecting anticoagulant delivery, or another internal device fault.

Clinical Use Tip

Never troubleshoot a PrismaFlex clotting alarm on an active patient without involving the clinical care team. Move or support the patient on a backup device or therapy plan first if extended troubleshooting is needed. Do not rely on a CRRT device with unresolved clotting, pressure, or anticoagulant delivery alarms.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PrismaFlex reported Alarm: HP Cartridge Clotted during CRRT treatment. Staff reported cartridge pressure drop alarm and concern for possible hemopurification cartridge clotting."

Cause

What was observed during troubleshooting.

Example:
"Inspection found restricted blood-flow tubing at the cartridge inlet. Line was corrected and alarm cleared. No device pressure measurement failure was identified during observation."

Resolution

What action was taken.

Example:
"Unclamped and rerouted restricted blood-flow line. Verified HP Cartridge Clotted alarm cleared and pressures stabilized. Device remained in service after clinical team confirmed treatment could continue."

Helpful Details to Include (If Known)

Final Thought

The HP Cartridge Clotted alarm should be treated as a therapy-impacting pressure condition until proven otherwise. Start with patient safety, confirm the exact alarm, check for line restriction, cartridge clotting, and anticoagulation-delivery issues, then escalate when pressure measurement or syringe pump failure is suspected. Clear documentation helps separate clinical circuit clotting from equipment failure and supports safe return to service.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide