On this page
Asset Type
Continuous Renal Replacement Therapy (CRRT)
Manufacturer
Model
What This Guide Helps With
This guide helps troubleshoot the Baxter PrismaFlex Plasmafilter Clotted alarm. This alarm may indicate high plasmafilter pressure drop, treatment interruption, suspected plasmafilter clotting, restricted blood flow, anticoagulation delivery issues, or unreliable pressure measurement during CRRT or TPE therapy.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
- Remove the PrismaFlex from active patient treatment if the alarm cannot be cleared safely or if filter clotting is suspected.
- Do not continue therapy through a clotted or partially clotted plasmafilter.
- Follow facility policy for returning blood, ending treatment, or changing the set.
- Verify that the patient is supported by clinical staff before any troubleshooting continues.
Expected: Patient therapy is protected, and the device is not used with a compromised extracorporeal circuit.
Why it matters: A clotted plasmafilter can affect therapy delivery, pressure accuracy, anticoagulation effectiveness, and patient safety.
2. Verify the Reported Alarm and Behavior
- Confirm that the displayed alarm is exactly: Alarm: Plasmafilter Clotted.
- Observe whether the alarm occurs during active treatment, after pressure trends gradually increase, after replacement flow or plasma-loss rate changes, after anticoagulant or citrate delivery changes, or intermittently with patient movement or line restriction.
- Look for associated symptoms such as high filter pressure drop, Plasmafilter Is Clotting advisory, TMPa Excessive Caution condition, visible clotting, restricted blood flow, repeated pressure alarms, or syringe pump/PBP delivery issues.
Expected: The alarm is reproducible or clearly tied to pressure trends or treatment conditions.
Why it matters: This confirms a true plasmafilter pressure-drop or clotting condition rather than a temporary setup or flow issue.
3. Check for Immediate Line Restrictions
- Inspect the blood-flow path for obvious restrictions.
- Check for clamped access or return lines, partially closed clamps, kinked tubing, compressed tubing, improper set routing, or patient positioning causing line restriction.
- Correct any visible restriction and observe whether pressures return toward normal.
Expected: Blood-flow lines are fully open, properly routed, and free of obstruction.
Why it matters: Clamped or kinked lines can create abnormal pressure conditions that mimic or accelerate plasmafilter clotting.
4. Assess the Plasmafilter and Circuit for Clotting
- Visually inspect the loaded set and plasmafilter as allowed by facility procedure.
- Look for darkened or streaked filter fibers, visible clots, increasing pressure trends, sluggish blood flow, repeated filter pressure-drop alarms, or a recent history of inadequate anticoagulation.
- If clotting is present or strongly suspected, change the set through STOP according to clinical protocol.
Expected: A clotted or compromised set is removed from patient use.
Why it matters: Once clotting is established, continued troubleshooting without changing the set may delay therapy and increase patient risk.
5. Review Anticoagulation Status With Clinical Staff
- Coordinate with the clinical team to review patient clotting parameters and anticoagulation status.
- Clinical checks may include the anticoagulation order, recent anticoagulation changes, ACT, aPTT, ionized calcium, facility-specific clotting values, citrate protocol status, or evidence of frequent circuit clotting.
Expected: Clinical staff determine whether anticoagulation is appropriate for the treatment.
Why it matters: Inadequate anticoagulation is a common real-world cause of plasmafilter clotting and repeated filter-related alarms.
6. Check Anticoagulant or Citrate Delivery
- From a Clinical Engineering perspective, verify whether the device is capable of delivering anticoagulant or citrate as intended.
- Check syringe pump operation, syringe installation, syringe size, syringe seating, PBP pump operation if citrate is delivered through PBP, pump door closure, tubing placement through the pump segment, and any related syringe pump, PBP, or flow alarms.
Expected: Anticoagulant or citrate delivery hardware appears properly loaded and operating without related alarms.
Why it matters: A failed syringe pump or PBP pump can cause inadequate anticoagulation delivery, leading to filter clotting.
7. Review Replacement Flow and Plasma-Loss Rate
- Check whether replacement flow or patient plasma-loss rate was recently increased or is excessive for the treatment setup.
- Work with clinical staff to reduce replacement flow and/or plasma-loss rate if clinically appropriate.
Expected: Flow settings are appropriate for the therapy, filter, and patient condition.
Why it matters: Excessive filtration or plasma-loss demand can increase filter stress and contribute to clotting or high pressure drop.
8. Check Pressure Measurement Reliability
- If the alarm occurs without visible clotting, line restriction, or clinical explanation, evaluate whether pressure measurement may be inaccurate.
- Check externally for pressure pod seating, incorrect tubing placement, moisture, debris, or damage near pressure-sensing areas, loose pressure components, or repeated pressure-related alarms across different sets.
Expected: Pressure monitoring components are clean, correctly positioned, and physically intact.
Why it matters: Pressure measurement failure can falsely report excessive pressure drop or clotting conditions.
9. Isolate Accessory, Clinical, or Device Cause
- If unclamping or correcting tubing clears the alarm, the issue was likely line restriction.
- If visible clotting is present, the plasmafilter or set is the likely cause and should be changed.
- If clotting repeats across sets, review anticoagulation, citrate delivery, and therapy settings.
- If delivery pumps fail or anticoagulant/citrate delivery is unreliable, remove the device for service evaluation.
- If pressure readings remain abnormal without a real circuit cause, suspect pressure sensing failure.
Expected: The cause is narrowed before assuming internal failure.
Why it matters: Plasmafilter Clotted alarms may be caused by clinical conditions, setup issues, therapy settings, pump delivery problems, or device pressure sensing faults.
10. Stop When Resolved
- If the alarm clears after correcting a clamp, kink, tubing restriction, flow setting, or pump loading issue, monitor the device for stable pressures before returning it to service.
- If therapy continues safely, document the correction and final device status.
Expected: The PrismaFlex operates without repeated Plasmafilter Clotted alarms.
Why it matters: A resolved alarm with stable pressure trends supports returning the device to clinical use.
If the Problem Persists
If external restrictions, set condition, anticoagulation status, flow settings, and loading issues have been ruled out, persistent Plasmafilter Clotted alarms may indicate repeated clinical clotting conditions or a device-related delivery or pressure measurement failure.
Remove the PrismaFlex from service if the alarm repeats with a new set, pressure readings appear inaccurate, syringe pump delivery is unreliable, PBP pump delivery is unreliable, anticoagulant or citrate delivery fails, or multiple pressure-related alarms occur without a clear circuit cause.
Label the device Out of Service and send it for repair or bench evaluation.
Bench evaluation may include syringe pump testing, PBP pump testing, pressure sensing verification, and related diagnostics. Replace the failed pump or pressure-sensing component if malfunction is confirmed.
Clinical Use Tip
Never troubleshoot on an active patient longer than necessary. Move the patient to a backup device, change the set, or transfer therapy according to clinical protocol before continuing troubleshooting if clotting is present or treatment delivery is compromised.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"PrismaFlex reported Alarm: Plasmafilter Clotted during treatment. Staff reported increasing filter pressure drop and therapy interruption."
Cause
What was observed during troubleshooting.
Example:
"Inspection found restricted blood-flow tubing from a partially closed clamp, causing elevated pressure drop across the plasmafilter. No device hardware failure confirmed."
Resolution
What action was taken.
Example:
"Corrected tubing restriction, verified clamps and line routing, confirmed pressure readings stabilized, and returned device to service after clinical staff verified treatment status."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
The Plasmafilter Clotted alarm should be treated as both a patient safety and therapy delivery concern. Start with the circuit, clamps, tubing, and treatment conditions before suspecting internal failure. If clotting is confirmed, change the set. If the alarm repeats without a clear external or clinical cause, remove the device from service and escalate for pump and pressure-sensing evaluation. Clear CCR documentation helps show whether the issue was circuit-related, clinical, or device-related.
That is successful troubleshooting.