On this page
Asset Type
Continuous Renal Replacement Therapy (CRRT)
Manufacturer
Model
What This Guide Helps With
This guide helps troubleshoot a Baxter PrismaFlex Return Extremely Positive alarm. This alarm may occur when return pressure is elevated due to a restricted return line, catheter obstruction, patient movement, excessive blood flow for the access condition, set-loading issue, pressure pod problem, or return pressure-sensing failure.
Because elevated return pressure can indicate restricted blood return to the patient or unsafe circuit pressure, this alarm should be treated as a meaningful clinical and equipment condition until the cause is identified.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Confirm the patient is being safely supported during the CRRT alarm condition.
- Notify the clinical team if therapy is interrupted or return pressure remains elevated.
- Follow facility policy for patient monitoring, CRRT escalation, blood return, treatment interruption, or transfer to another CRRT device.
- Do not ignore repeated return pressure alarms during active therapy.
Verify the Reported Alarm and Behavior
- Confirm the PrismaFlex is displaying Alarm: Return Extremely Positive.
- Observe whether the alarm occurs during startup, prime, shortly after treatment begins, or during active treatment.
- Check whether the alarm appears only during patient movement, coughing, suctioning, or repositioning.
- Document whether the alarm self-clears within the allowed self-clear interval or repeatedly returns after clearing.
Allow Self-Clearing When Clinically Appropriate
- If the patient is stable and the alarm appears related to a brief event, allow the machine’s self-clearing attempt when appropriate.
- Watch for temporary triggers such as patient movement, coughing, suctioning, repositioning, or short-term catheter pressure changes.
- If the alarm clears and does not recur, continue monitoring the treatment and document the event.
Inspect the Return Line for External Obstruction
- Trace the return line from the PrismaFlex set to the patient connection.
- Check for closed clamps, partially closed clamps, kinked tubing, sharp bends, or line tension.
- Look for tubing pinched under the machine, bed rail, blanket, or patient.
- Correct any visible obstruction and verify that return pressure stabilizes.
Check Catheter Position and Patient-Related Causes
- Assess whether the return catheter may be restricted, clotted, malpositioned, or affected by patient position.
- Coordinate with nursing or the clinical team to reposition the patient if appropriate.
- Confirm there is no catheter tension, pulling, or incorrect lumen connection.
- Allow clinical staff to flush, reposition, or evaluate the catheter according to hospital protocol.
- Do not perform catheter interventions outside Clinical Engineering scope.
Review Blood Flow Settings
- Check whether the blood flow rate is set higher than the catheter or patient access can tolerate.
- If clinically appropriate, have the care team reduce blood flow and observe the return pressure response.
- If return pressure decreases after blood flow is lowered, document the finding and confirm the treatment plan with the clinical team.
Use RELEASE CLAMP if Available and Appropriate
- If the PrismaFlex provides a RELEASE CLAMP option for this condition, follow facility procedure and manufacturer guidance.
- Use the option only when appropriate to relieve excess return-line pressure during troubleshooting or treatment management.
- Do not bypass safety alarms or repeatedly release pressure without identifying the cause.
Check for Accessory or Set-Related Issues
- Inspect the loaded PrismaFlex set for visible problems.
- Check for an improperly seated return pressure pod.
- Look for a wet, contaminated, or damaged pressure pod interface.
- Verify tubing is loaded correctly and is not pulled tight across the pressure monitoring area.
- Inspect for visible clotting in the return path, set damage, or abnormal tubing collapse.
- If the set appears compromised or clotting is suspected, follow clinical procedure for ending treatment or changing the set.
Determine Whether the Issue Is Patient, Circuit, or Device Related
- If the alarm follows coughing, movement, suctioning, or position changes, suspect a transient patient or access-related pressure change.
- If the alarm improves after removing a kink or opening a clamp, suspect return-line obstruction.
- If the alarm improves after catheter flush or repositioning by clinical staff, suspect catheter restriction or malposition.
- If the alarm improves after lowering blood flow, suspect access limitation or excessive flow for the current catheter condition.
- If the alarm persists with no obstruction, acceptable access, and proper setup, suspect return pressure sensing failure or an internal device issue.
Stop When Resolved
- If the alarm clears after correcting a line, catheter, flow, or setup issue, confirm return pressure remains stable.
- Confirm no related alarms are active.
- Verify the circuit is operating normally.
- Return the device to service if clinically appropriate.
- Document the cause and resolution clearly.
If the Problem Persists
If the Return Extremely Positive alarm continues after checking patient movement, return line obstruction, catheter condition, blood flow settings, set loading, and pressure pod condition, external and common causes have been ruled out.
Persistent Return Extremely Positive alarms may indicate failed return pressure sensing, incorrect return pressure calibration, a pressure pod detection problem, an internal pressure monitoring fault, or a circuit condition requiring treatment termination.
Remove the device from service if the alarm cannot be safely resolved. Label it Out of Service and send it for repair or bench evaluation.
For service evaluation, check and calibrate the return pressure sensor according to approved service procedures. Replace the return pressure sensor or related hardware if readings remain inaccurate.
Clinical Use Tip
Never troubleshoot unresolved CRRT pressure alarms on an active patient longer than clinically appropriate. Move the patient to a backup CRRT device or alternate treatment plan first if safe therapy cannot be maintained.
Never rely on a PrismaFlex with repeated unexplained return pressure alarms.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"PrismaFlex reported Return Extremely Positive alarm during treatment. Return pressure exceeded the user-set warning limit and the alarm recurred after an initial self-clear attempt."
Cause
What was observed during troubleshooting.
Example:
"Return line was found kinked near the patient connection, causing excessive return pressure. Pressure stabilized after tubing was rerouted."
Resolution
What action was taken.
Example:
"Corrected return-line routing, verified clamps were open, observed stable return pressure, confirmed no related alarms were active, and returned device to service."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Return Extremely Positive alarms should be treated as meaningful pressure events until proven otherwise. Start with patient safety, verify the alarm, check simple external causes, and work logically toward catheter, circuit, or sensing issues. If the alarm persists after common causes are ruled out, remove the device from service and escalate for repair. Clear CCR documentation protects the patient, the clinical team, and the equipment history.
That is successful troubleshooting.