Baxter PrismaFlex

Alarm: Scale Open

On this page

Asset Type

Continuous Renal Replacement Therapy (CRRT)

Manufacturer

Baxter

Model

PrismaFlex

What This Guide Helps With

This guide helps troubleshoot the Scale Open alarm during PrismaFlex priming when a scale cannot fully close or the machine cannot detect the closed scale condition.

This alarm prevents setup from continuing until the scale, bag, carrying bar, and closure detection are correct.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check for Obstruction Around the Scale

Verify Bag Positioning

Check the Carrying Bar and Bar Tray

Confirm Carrying-Bar Handle Orientation

Confirm the Scale Locks Closed

Isolate Whether the Issue Is Mechanical or Sensor-Related

Check Scale Sensor Status in Diagnostics

Stop When Resolved

If the Problem Persists

If the scale is correctly loaded, unobstructed, aligned, latched closed, and the carrying-bar handle is rotated down, but the PrismaFlex still reports Scale Open, external and common setup causes have been ruled out.

A persistent Scale Open alarm may indicate a scale closure sensor, latch, wiring, or scale assembly failure.

Remove the PrismaFlex from service, label it Out of Service, and send it for repair or bench evaluation. If the scale sensor fails testing in Diagnose Screen – Scale Diagnose, replace the affected scale assembly according to service procedure.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first.

Do not continue priming or connect a patient to the PrismaFlex while a Scale Open alarm is unresolved. Use another CRRT system if therapy needs to begin before the machine is corrected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PrismaFlex reported Alarm: Scale Open during priming. Staff unable to continue setup because the scale would not close or the machine did not detect closure."

Cause

What was observed during troubleshooting.

Example:
"Scale bag and carrying bar area inspected. Bag position and carrying-bar alignment corrected. Carrying-bar handle was not fully rotated down toward the floor, preventing proper scale closure."

Resolution

What action was taken.

Example:
"Repositioned bag, centered carrying bar, rotated handle downward, pushed scale closed until locked, and pressed CONTINUE. Alarm cleared and priming resumed normally. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

A Scale Open alarm should be approached as a setup and safety issue first. Most causes are external: bag position, obstruction, carrying-bar alignment, or handle orientation. If those are corrected and the machine still cannot detect closure, escalation for scale sensor or assembly repair is appropriate. Clear documentation protects the patient, the clinical team, and the repair process.

That is successful troubleshooting.